3.1
Installation of Components
3.1.1 How to Attach the Biopsy Guidance Module
You can install the Biopsy Guidance Module with the Selenia Dimensions power on or off.
Caution:
To prevent damage or alignment problems with the
Needle Guidance Stage, be careful when you move the
Biopsy Guidance Module.
Caution:
The Affirm Biopsy Guidance Module weighs 15 pounds.
When you move it, be sure to have a secure grip on the
handles.
Figure 7:
Installation of the Biopsy Guidance Module
1.
Move the Compression Device down.
2.
Hold the Biopsy Guidance Module by both handles.
3.
Slide the top hooks (item 2) of the Biopsy Guidance Module into the slots marked
with the Affirm icon (item 2A) on the front of the Selenia Dimensions gantry. Make
sure the top and bottom hooks attach to the Selenia Dimensions gantry.
4.
Push the Lock Levers (item 3) on the Biopsy Guidance Module into the Up position to
lock the Module against the Selenia Dimensions.
5.
Align the red dot on the cable (item 4) from the Biopsy Guidance Module with the red
dot on the receptacle on the Selenia Dimensions gantry. Connect the cable to the
receptacle.
Chapter 3:
Содержание Affirm
Страница 1: ......
Страница 2: ......
Страница 3: ...User Guide For Software Version 1 7 Part Number MAN 03209 Revision 005 January 2015...
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 16: ......
Страница 24: ......
Страница 32: ......
Страница 54: ......
Страница 58: ......
Страница 62: ......
Страница 64: ......
Страница 66: ......
Страница 69: ......
Страница 70: ......