A.1 Affirm Measurements
A. Height
37.1 cm (14.6 inches)
B. Width
37.8 cm (14.9 inches)
C. Depth
35.6 cm (14 inches)
A.2 Biopsy Guidance Module
Weight
15 pounds
Accuracy
± 1 mm
Range of Movement
X-axis: ±35 mm
Y-axis: +72.8 mm
Z-axis: + 161 mm
Speed of Motorized Movements
Continuous: No faster than 5 mm per second)
Incremental: 0.5 mm steps
Power System
Input from Selenia Dimensions:
+15Vdc±10% and +5Vdc±10%
Output: +12Vdc
Note
The Affirm biopsy guidance system is designed for + 1 mm
targeting precision when a biopsy device is mounted to the
system with an applicable device holder.
A.3 Biopsy Control Module
Display Window
Touch screen controls
Weight
3 pounds
Power System
Input from Biopsy Guidance Module: +5Vdc±10%