Appendix B
–
Communications 19
HP-310
Command
Description/Allowed Data/Response
FM
Rate Units
“0”
for seconds
“1”
for minutes
“2”
for hours
“3”
for days
“FLOW UNITS=(DATA)”
(DATA)
shall be:
“SEC”
for seconds
“MIN”
for minutes
“HR
”
for hours
“DAY”
for days
NB
Max Sample Time
“1”
to
“80”
“MAX M TIME=(DATA)”
LF
Out Low
“0.000”
to a maximum value of the Out High setting
“4mA FLOW =(DATA)”
AF
Out High
Minimum is the Out Low Setting (LF) to a maximum of the
following:
“99999.999”
if RD = 3
“999999.99”
if RD = 2
“9999999.9”
if RD = 1
“ 99999999”
if RD = 0
“20mA FLOW =(DATA)”
PA
Password
“0”
to
“9999”
“PASS WORD =(DATA)”
OC
Current Out
“0”
- Current output follows rate.
“1”
- Current output set to 4mA.
“2”
- Current output set to 12mA.
“3”
- Current output set to 20mA.
For
“0”
, response =
“ Output equal to input.”
For
“1”
, response =
“ Output is 4mA.”
For
“2”
, response =
“ Output is 12mA.”
For
“3”
, response =
“ Output is 20mA.”