Principle of Operation 9
HP-310
3-2
System Response Time
The analog output response time to reach steady state due to a change in
the flow rate is approximately (1/8) seconds. When flow stops, the time
for the analog output to return to 4 mA will be between 3 and 12
seconds, depending on the Maximum Sample Time (MST) setting.
MST is adjusted using the NB= (DATA) command, where NB is a value
between 1 and 80. The default MST setting is NB= 1. Adjusting the
MST is only recommended for low flow applications where the
minimum input frequency is below 1 Hz.