![HOFFEN BPM-8417 Скачать руководство пользователя страница 63](http://html1.mh-extra.com/html/hoffen/bpm-8417/bpm-8417_instruction-manual_2143753063.webp)
3
63
IM_POJM180417 ARM BLOOD PRESSURE MONITORBASIC 2018-07-25
Table 2
Enclosure Port
Phenomenon
Basic EMC
standard
Immunity test levels
Home Healthcare
Environment
Electrostatic
Discharge
IEC 61000-4-2
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
Radiated RF EM field
IEC 61000-4-3
10V/m
80MHz-2.7GHz
80% AM at 1kHz
Proximity fields from RF wireless
communications equipment
IEC 61000-4-3
Refer to table 3
Rated power frequency magnetic
fields
IEC 61000-4-8
30A/m
50Hz or 60Hz