p4
2. Important information
• The safety lid must be in place before connecting
the power leads to a power supply.
• Turn all power supply controls off and disconnect
the power leads before removing the safety lid.
• Circulate only water or 50/50 water/ethylene glycol
through the heat exchanger. Never introduce anti-
freeze or any organic solvent into any part of the
instrument. Organic solvents will cause irreparable
damage to the unit!
• Do not connect the heat exchanger to a water tap
or any coolant source where the water pressure is
unregulated.
• Do not operate with buffer temperature above 45
°C. All plastic parts are rated for 45 °C continuous
duty. Circulate coolant through the heat exchanger
during electrophoresis to minimize heating. Addi-
tional passive cooling actions include chilling the
buffer before use, running the unit in a cold room,
or both. Overheating will cause irreparable damage
to the unit!
• If running only one gel, block off the unused part
of the core with a glass plate. Do not fill this side
with buffer.
• Only accessories and parts approved or supplied by
Hoefer, Inc. may be used for operating, maintaining,
and servicing this product.