background image

RXW EC-5 and 10H Soil Moisture Sensor (RXW-SMx-xxx) Manual 

1-800-LOGGERS 2 

www.onsetcomp.com 

Specifications (continued) 

Memory 

16 MB 

Dimensions 

RXW-SMC-xxx soil probe: 89 x 15 x 1.5 mm (3.5 x 0.62 x 0.06 in.) 
RXW-SMD-xxx soil probe: 160 x 32 x 2 mm (6.5 x 1.25 x 0.08 in.) 
Cable length: 5 m (16.4 ft) 
Mote: 16.2 x 8.59 x 4.14 cm (6.38 x 3.38 x 1.63 inches) 

Weight 

RXW-SMC-xxx sensor and cable: 180 grams (6.3 oz) 
RXW-SMD-xxx sensor and cable: 190 grams (6.7 oz) 
Mote: 223 g (7.87 oz) 

Materials 

Sensor: Weatherproof 
Mote: PCPBT, silicone rubber seal 

Environmental Rating 

Mote: IP67, NEMA 6 

Compliance Marks 

RXW-SMC-900 and RXW-SMD-900: See last page 

 

RXW-SMC-868 and RXW-SMD-868: The CE Marking identifies 
this product as complying with all relevant directives in the 
European Union (EU). 

RXW-SMC-922 and RXW-SMD-922: See last page 

Note 1:  (RXW-SMC-xxx and RXW-SMD-xxx) The sensor is capable of providing readings outside the standard volumetric 

water content range. This is helpful in diagnosing sensor operation and installation. See the 

Sensor

 

Operation

 

section for more details. 

Note 2:  (RXW-SMC-xxx) This is a system level accuracy specification and is comprised of the probe’s accuracy of ±0.03 

m³/m³ typical (±0.02 m³/m³ soil specific) plus the mote accuracy of ±0.001 m³/m³ at 25°C (77°F). There are 
additional temperature accuracy deviations of ±0.003 m³/m³ / °C maximum for the probe across operating 
temperature environment, typical <0.001 m³/m³ / °C. (The temperature dependence of the mote is negligible.) 

Note 3:  (RXW-SMC-xxx) Given the nature of the sensor design and sensor operating frequency, the system has inherent 

susceptibilities to Radio Frequency signals. The accuracy specification when subjected to certain RFI 
environments, such as those outlined in IEC 61000-4-3 and IEC 61000-4-6, is reduced to 0.061 m³/m³. 

The system level accuracy will be particularly affected when placed in an electric field of 3 V/m or greater in the 
70 MHz range. RFI mitigation practices and physical deployment changes may reduce the systems susceptibility. 

Note 4:  (RXW-SMD-xxx) This is a system level accuracy specification and is comprised of the probe’s accuracy of ±0.03 

m³/m³ typical (±0.02 m³/m³ soil specific) plus the mote accuracy of ±0.003 m³/m³ at 25°C (77°F). There are 
additional temperature accuracy deviations of ±0.003 m³/m³ / °C maximum for the probe across operating 
temperature environment, typical <0.001 m³/m³ / °C. (The temperature dependence of the mote is negligible.) 

Note 5:  (RXW-SMD-xxx): Given the nature of the sensor design and sensor operating frequency, the system has inherent 

susceptibilities to Radio Frequency signals. The accuracy specification when subjected to certain RFI 
environments, such as those outlined in IEC 61000-4-3 and IEC 61000-4-6, may be significantly reduced. 

The system level accuracy will be particularly affected when placed in an electric field of 3 V/m or greater in the 
150 KHz to 1000 MHz range. RFI mitigation practices and physical deployment changes may reduce the systems 
susceptibility, however will yield reduced system accuracy. If deployments are planned in high RFI energy 
environments, Onset recommends on-site testing to determine system level accuracy. 

 

Mote Components and Operation 

 

 

Sensor Mote Closed, Front 

Sensor Mote Closed, Back 

 

 

Sensor Mote Opened 

Mounting Tab:

 Use the tabs at the top and bottom of the mote 

to mount it (see 

Deploying and Mounting

). 

Solar Panel:

 Position the solar panel towards the sun to charge 

the mote batteries (see 

Deploying and Mounting

). 

Mounting 
Tab 

Solar Panel 

Solar Panel Cable 

Button 

USB Port 

Battery Holder 

LCD Screen 

Latch 

Eyelet 

Ground Wire 
Port 

Sensor Cable 

Antenna 

LEDs 

Содержание RXW 10HS

Страница 1: ...d cable can safely operate at below freezing temperatures to 40 C F the soil moisture data collected at these extreme temperatures is outside of the sensor s accurate measurement range Extended temperatures above 50 C 122 F will decrease mote battery life Wireless Mote Operating Temperature Range 25 to 60 C 13 to 140 F with rechargeable batteries 40 to 70 C 40 to 158 F with lithium batteries Radio...

Страница 2: ... as those outlined in IEC 61000 4 3 and IEC 61000 4 6 is reduced to 0 061 m m The system level accuracy will be particularly affected when placed in an electric field of 3 V m or greater in the 70 MHz range RFI mitigation practices and physical deployment changes may reduce the systems susceptibility Note 4 RXW SMD xxx This is a system level accuracy specification and is comprised of the probe s a...

Страница 3: ...blink and then the bars in the icon will cycle from left to right The x icon will blink during the last step in the network registration process see Adding the Mote to the RX Wireless Sensor Network for details LCD Symbol Description This indicates a problem with the sensor itself the mote is operational Check the sensor and make any adjustments to it as needed Contact Onset Technical Support if t...

Страница 4: ... the next time the RX3000 station connects to HOBOlink See the HOBOlink Help for details on how to change the logging and connection intervals view data check mote status add the mote to a map and more Deployment and Mounting Mounting and Positioning the Mote Mount the mote to a mast or pipe using cable ties or affix the mote to a wooden post or flat surface with screws Insert the cable ties or sc...

Страница 5: ...soil auger to make a hole to the desired depth an angled hole is best and push the probes into undisturbed soil at the bottom of the hole Alternatively dig a hole and push the probes into the side of the hole If the probe has a protective cap on the end remove it before placing the probe into the hole To push the probe into the soil use a PVC pipe with slots for the sensor and a longer slot for th...

Страница 6: ...attery recharges to 2 3 V If the batteries are unable to be recharged replace them with fresh rechargeable batteries Note if you install used rechargeable batteries that together are less than 2 3 V the mote will not resume operation To replace rechargeable batteries 1 Open the mote door 2 Remove the old batteries and install fresh ones observing polarity 3 Make sure the solar panel cable is plugg...

Страница 7: ...ning the equipment off and on the user is encouraged to try to correct the interference by one of the following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help This d...

Отзывы: