background image

Warranty

Onset Computer Corporation (“Onset”)

warrants to the original end-user purchaser for

a period of one year from the date of original

purchase that the HOBO

®

 product(s)

purchased will be free from defect in material

and workmanship. During the warranty period

Onset will, at its option, either repair or

replace products that prove to be defective in

material or workmanship. This warranty shall

terminate and be of no further effect at the

time the product is (1) damaged by extraneous

cause such as fire, water, lightning, etc. or not

maintained in accordance with the

accompanying documentation; (2) modified; (3)

improperly installed; (4) repaired by someone

other than Onset; or (5) used in a manner or

purpose for which the product was not

intended.

Returning Products to Onset

Direct all warranty claims to place of

purchase. Before returning a failed unit, you

must obtain a Return Merchandise

Authorization (RMA) number from Onset. You

must provide proof that you purchased

the Onset product(s) directly from Onset

(purchase order number or Onset invoice

number). Onset will issue an RMA

number that is valid for 30 days. You

must ship the product(s), properly

packaged against further damage, to

Onset (at your expense) with the RMA

number marked clearly on the outside of

the package. Onset is not responsible for

any package that is returned without a

valid RMA number or for the loss of the

package by any shipping company. Loggers

must be clean and free of any toxins before

they are sent back to Onset or they may be

returned to you.

Repair Policy

Products that are returned after the warranty

period or that are damaged by the customer as

specified in the warranty provisions can be

returned to Onset with a valid RMA number

for evaluation.

Please contact Onset for more information and

prices on:

ASAP Repair Policy

Onset will expedite the repair of a returned

product.

Data-back

 Service

HOBO

®

 data loggers store data in nonvolatile

EEPROM memory. Onset will, if possible,

recover your data to a disk.

Tune Up

 Service

Onset will examine and retest any HOBO

®

data logger.

Thank you for buying a HOBO H8 Pro Temp/RH IS data logger. With proper care

it will give you years of accurate and reliable measurements.
This manual covers the HOBO H8 Pro Temp/RH IS product. This logger stores up

to 65,291 time-stamped measurements.
Unlike most other HOBOs, the HOBO Pro does not have the wrap-around-when-

full option for storing data; its large memory capacity eliminates the need for this

function in most cases.

Safety Information  - Read First

WARNING

: The H08-032-IS logger is rated as Intrinsically Safe, however,

available communication accessories are not. Do not attempt to download or

relaunch the logger using the HOBO Shuttle, handheld PDA or any host computer

in a hazardous environment, as these devices do not carry the Intrinsically Safe

rating and are not intrinsically compatible with HOBO-IS Loggers. See the “Logger

Specifications” for the complete Intrinsically Safe listing and see the section of this

manual entitled “Intrinsically Safe” for more information about hazardous location

and the H08-032-IS.

WARNING:

  Fire, Explosion, and Severe Burn Hazard. The logger contains a

lithium battery. The battery may explode if the logger is exposed to extreme heat or

conditions that could damage or destroy the battery case. Do not attempt to

recharge or heat the logger or battery above +185

o

F (+85

o

C). Do not dispose of the

logger or battery in fire. Do not expose the contents of the battery to water. Dispose

of the battery according to local regulations for lithium batteries.

WARNING:

  Only qualified personnel should install and service equipment

located in Hazardous (Classified) Areas.

Common Specifications

Operating range (logger): -30°C to +50°C (-22°F to +122°F),

0 - 100% RH, HOBO Pro RH/Temp IS should be mounted

so that water does not impact or collect in the RH sensor.

RH Sensor operating environment: 0°C to +50°C (+32°F to

+122°F) in intermittent condensing environments up to
+30°C; and above +30°C in non-condensing environments.
Sensor requires protection from rain, splashing, mist,
dust, and airborne chemicals such as salt and ammonia.

Time accuracy: approx. ±1 minute per week (±100 ppm at

+20°C or +68°F), full dependence shown in Plot A.

Measurement capacity: 65,291 standard-resolution (8-bit) measurements, 32,645

high-resolution (12-bit) measurements or 21,763 measurements if one channel

uses standard-resolution and the other channel uses high-resolution. RH

measurements use standard resolution only. All measurements are stored in

nonvolatile memory, with seven levels of data archiving (See Non-Volatile

Memory).

Data offload time: 1 minute typical

Size: 4.0" H x 3.2" W x 2.0" D

Weight: approx. 3.7 oz.

Battery: Tadiran TL-5902, lithium, (3.6V 1.2Ah), user-replaceable (Use only Onset

part # HP-B)

Battery life (continuous use): 3 years
Storage temperature: -30°C to +75°C (-22°F to +167°F)

Intrinsically Safe Rating:  IS, Class I, II, Division 1, Groups A-G, Temperature

Code T4 (

<135

o

C); NI, Class I, Division 2, Groups A-D, Temperature Code T4

(

<135

o

C); S, Class II, Division 2, Groups F and G

Note: Before replacing the back plate, check that the logger seals (O-ring, screw, jack

cap) are not contaminated by dust, dirt or other materials. All components must be

cleaned of any contaminants to ensure a weathertight seal. Make sure the back plate

O-ring has a light film of lubricant (Dow Corning DC 111 or Nye Lubricants

NYOGEL). Insert the O-ring in the groove and replace the back plate using the screw

with O-ring. The screw should be tightened until it is snug, be careful not to

overtighten. The screw will go in smoothly and suddenly tighten; rotate the screw no

more than 1mm more. Do not use lubricant on jack cap.

Intrinsically Safe

Factory Mutual Research has certified the H08-032-IS is safe for use in certain

hazardous areas. The National Electrical Code has classified many types of

hazardous locations. The H08-032-IS has been approved as intrinsically safe in

Class I and II locations which are areas where ignitable concentrations of gas

(Class I) and/or dust (Class II) exist. The NEC further classifies hazardous location

by Division (1 or 2). The division designation refers to the likelihood that the area

will contain ignitable concentrations of gas or dust. Division 1 locations will have

ignitable concentrations of gas or dust continuously or under normal operation

conditions. Division 2 locations will have ignitable concentrations of gas or dust

only during infrequent or abnormal operations (which may include areas located

near Division 1 locations). The H08-032-IS has been approved for use in both

Division 1 and 2 locations. The hazardous area classification scheme is further

defined by listing the types of ignitable gases and dusts that may be present in a

hazardous location. Ignitable gases and dusts are sorted into groups with similar

ignition characteristics. The H08-032-IS is certified for all possible groups:  A-G for

gas and A-D, F and G for dust. Finally every Intrinsically Safe Device is given a

temperature rating. The H08-032-IS has a T4 rating, which means that no part of
the device will become warmer than +275

o

F (+135

o

C) in a worst-case fault. For more

information on Hazardous (Classified) Locations visit the Underwriters Laboratory

web site at www.ul.com/hazloc/define.htm.

Non-Volatile Memory

The HOBO Pro uses a high-capacity Flash EEPROM to store data. This storage is

non-volatile and will retain the data even if the battery is removed. To save power

and maximize the life of the flash memory device, 32 bytes of data are buffered in

RAM prior to writing to the flash memory.  In the case of a dead battery, or the

unlikely event that power is interrupted during logging, the data in this RAM buffer

will be lost. For example, if the sample interval was set to one half hour and one

channel with low resolution, then up to 16 hours of data could be lost.

NOTE: The logger software's absolute humidity calculation does not use an actual

pressure measurement but assumes an ambient pressure of 1 atmosphere (14.7 PSI.)

Service and Support

HOBO

®

 products are easy to use and

reliable. In the unlikely event that you have

a problem with the hardware or software,

please read the following.

Who do I contact?

Contact the company that you bought the

loggers from: Onset Computer Corporation

or an Onset Authorized Dealer.

Before calling, you can evaluate and often

solve your problem if you try the following:

1. Read this manual and the ReadMe file on

the software disk. It may only take a few

moments to get the answers you need.

2. Write down the events that led to the

problem. Have you changed anything in

your computer recently?  Are you doing

anything differently?
When contacting Onset Computer

Corporation, please indicate that you need

Technical Support for HOBO

®

 products.

Be prepared to:

1. Provide the product number which is

found on the side of the logger, the software

version and serial number if present on the

disk.

2. Provide details on the hardware and

software configuration of your computer

including: manufacturer, model number,

peripherals, and version of operating

system.

3. Completely describe the problem or

question. The more information you provide,

the faster and more accurately we will be

able to respond.

NOTE: Onset provides technical support to

one person for each software license.

Onset Technical Support

Onset Computer Corporation

470 MacArthur Blvd.

Bourne, MA 02532

Mailing:  PO Box 3450

Pocasset, MA 02559-3450

1-800-LOGGERS (1-800-564-4377)

Phone:  (508) 759-9500

Fax:  (508) 759-9100

E-mail:  [email protected]

www.onsetcomp.com