www.HMN.dk
HMN a/s
26
Trouble Shooting/guide for electrical issues with M2 MULTI
-
TIP chairs.
NB! AS A FIRST COURSE OF ACTION
CONCERNING ELECTRICAL ISSUES, AL-
WAYS TRY TO CHARGE THE CHAIR
FOR 24 HOURS!
1. If only the left/last battery light is on.
The chair is low on power and must be charged.
The chair will not elevate if only 1 battery light is
on.
2. If the 2 smaller lights at the arrow buttons on
the hand control is on.
The emergency STOP has been activated. It must
be deactivated by holding down the emergency
stop button on the hand control until the chair
beeps once and the 2 small diodes at the arrow
keys go out.
3. All 4 lights at the top of the hand control is
blinking.
The chair's control box has received an error sig-
nal. The chair must be reset. This is done by first
pressing the emergency stop button on the hand
control until the emergency stop is activated, and
then holding the button next to the emergency
stop button. First the chair will beep 3
-
4 times,
then it will lower the chair completely. When the
chair is stationary, it is reset.
If the chair after this is still blinking in all 4 top
lights on the hand control it means there is a
defect component.
Contact HMN for further advise.
Содержание Multi-Flexi-Tip 310810-B
Страница 8: ...HMN a s 7 ...
Страница 15: ...www HMN dk HMN a s 14 ...
Страница 16: ...HMN a s 15 ...
Страница 30: ...Midtager 22 2605 Brøndby Danmark Tlf 45 4396 2400 E mail info hmn dk Web www hmn dk HMN a s V 1 0 ...