4
Step
Operation
Displaying
Descriptions
1
Press
*
Press
→
CAL SP
CAL 00
Indicator in calibration mode
Enter to set ZERO
2
Press
*
------
3000
Set ZERO
After several seconds, indicator displays maximum weight
3
Load max weight
Press
*
3000
Display weight
4
Press
→
Press
↑
Or
→
0
←
000000
002000
If the weight is not the max weight, press
→
and
↑
or
→
0
←
to input actually weight
e.g.: weight is 2000
5
Press
*
------
2000
Calibrate
After several seconds, indicator displays weight, then finish
calibrate
§7. ADC Counts Display
Confirm load cell connect is correct to check ADC counts. After set the short pin, ADC counts should
around 100000-200000. After above setting, check ADC counts directly:
Step
Operation
Displaying
Descriptions
1
Press
*
CAL SP
Indicator in calibration mode
2
Press
*
-SEt-
Indicator in configuration mode
3
Press
*
-A -d-
ADC counts display
4
Press
→
43125
Enter to check ADC counts, and display 43125
5
Press
*
0
Return to weighing mode
§8. Factory Set
Step
Operation
Displaying
Descriptions
1
Press
*
CAL SP
Indicator in calibration mode
2
Press
*
-SEt-
Indicator in configuration mode
3
Press
*
-A -d-
Indicator in ADC counts display mode
4
Press
*
FACt
Select Back to factory set
5
Press
→
0
Back to factory set and return to weighing mode
Factory default: d=0.01
FS=150.00
FLt=2
AUtP=11
Adr=00
b=2400
§9. Weighing Unit
Step
Operation
Displaying
Descriptions
1
Press
*
On
000000
999999
0
Press
*
and
ON
to switch on indicator
If it displays [UE=21.X], user can set own unit
2
Press
↑
2.0000
Select unit
3
Press
→
000000
Digit flash
4
Press
↑
→
0
←
01.6535
Change digit number
If user unit is Hong Kong kg, enter 1.6535
5
Press
*
HE 0
Decimal/H
exadecimal
(lb And UU)
0---Decimal
1---
Hexadecimal
Содержание X1FM