4 - 29
Following is the relation between delay time and column flush.
(1)
With AUTO STOP at YES
(a)
When YES is selected for COLUMN FLUSH
∗
“n” indicates number of analyses specified with autosampler, etc.
Stabilizing time is the time from the end of one analysis until the start of the
next analysis.
Column flushing starts when delay time has ended.
(b)
When NO is selected for COLUMN FLUSH
∗
“n” indicates number of analyses specified with autosampler, etc.
Stabilizing time is the time from the end of one analysis until the start of the
next analysis. The pump stops when delay time has ended.
(2)
With AUTO STOP at NO
∗
“n” indicates number of analyses specified with autosampler, etc.
The pump remains in liquid delivery status.
Analysis 1 Analysis 2 Analysis 3
Analysis n
∗
Delay time
Stabilizing time
Pump stop
Delay time
Column flush
Gradient end
Pump remains in
liquid delivery status.
Analysis 1 Analysis 2 Analysis 3
Stabilizing time
Analysis n
∗
Analysis 1 Analysis 2 Analysis 3
Delay time
COLUMN FLUSH
Stabilizing time
Pump stop
Analysis n
∗