4 - 3
4.1.2
Programming Operation
This section discusses the programming of the FLOW mode.
An example is given where upper pressure limit is 20.0 MPa,
lower pressure limit is 0 MPa, analysis time is 0 min, flow rate is
1.0 mL/ min.
PROGRAM No.
=
1
PRESSURE VALUE
:
MAX
MIN
MPa
20.0
0.0
TIME FLOW
EVENT
(min) (ml/min)
1 2
3
4
0.0
1.000
12
22
32
42
Monitor screen
If manual setting screen is displayed, press INITIAL to
obtain the programming mode.
PROG
Program generation is started.
SET PROGRAM:
No.
[--]
<1-9> 1
Input program No. Valid program numbers are from 1 to
9.
MODIFY=0,
NEW=1 [--]
0
Generates a new program or modifies an existing
program.