-12-
Q1E-EP1457
3.3
Needle Guide Bracket EZU-PA7C2
1)
Confirm that the needle guide bracket is cleaned and disinfected,
and the probe is cleaned and disinfected/ sterilized.
2)
Fit the needle guide bracket to the probe following “3.3.1 Fitting
the needle guide bracket to the probe”. Never apply excessive force
to attach the needle guide bracket to the wrong position. It may
cause a hazard due to unstable biopsy.
3)
Fit the CIV-Flex ™ cover to the probe following “3.3.2 Placing the
probe and bracket into a transducer cover”.
4)
Attach the needle guide to the bracket following “3.3.3 Attaching
the needle guide to the bracket”.
5)
Fit the needle insert corresponding to the gauge number of the needle
to be used to the needle guide following “3.3.3 Attaching the needle
guide to the bracket”.
6)
Confirm that the needle guide angle corresponds to the angle
indicated on the monitor. The needle guide angle is engraved on
bracket following “3.3.4 Setting the needle guide angle”.
7)
In the case of the CIVCO Ultra Pro II, when the needle is released
quickly from probe, press the tab on the needle insert toward the
bracket. Move bracket and needle guide away from the needle. (See
Fig. 10)
8)
After the use of the needle guide bracket, it should be cleaned and
disinfected, then store it in an adequate place.
Tab on the
needle insert
Push
Needle
Fig. 10 Release the needle
Содержание C22P
Страница 48: ... 42 Q1E EP1457 Unit mm Fig 23 Dimensions ...
Страница 49: ... 43 Q1E EP1457 Unit mm Fig 24 External view with the attachment ...
Страница 50: ... 44 Q1E EP1457 Unit mm Fig 25 Dimensions with the Needle Guide Bracket ...
Страница 51: ... 45 E Q1E EP1457 Fig 26 External view with Ultra Pro II Needle Guide cover kit Unit mm ...