-32-
Q1E-EP1457
Prepare following items before manual cleaning and disinfection:
a)
Detergent: Cidezyme® (Johnson & Johnson, #2258) or another cleaning
agent with approved material compatibility for this medical device.
b)
Disinfectant: Cidex® OPA (Johnson & Johnson, # 20391) or another
disinfectant with approved material compatibility for this medical
device
c)
Cleaning brushes if applicable
d)
Two tanks, one for cleaning and one for disinfection - optional:
1 additional tank for rinsing with deionized/tap water (sufficient
size for immersion of the submergible part of the probe at full
length)
e)
Soft, fluff free cloth or single use towel
f)
50 ml syringe
g)
Personal protective equipment (gloves, water repellent protective
skirt, face protection mask or protective glasses, see also
instructions of the manufacturer for the detergent and the
disinfectant)
Manual Cleaning:
Prepare the detergent solution in a tank with cold water (please follow
the instructions of the detergent manufacturer regarding application,
dilution and contact time).
1)
The temperature of the detergent solution should be between 15-30 °C,
concentration is 1.6%. Please note the minimum contact time of the
detergent in the manufacturer’s instruction. If a differing
detergent is used, please also note the approved material
compatibility for the medical device.
2)
Separate the needle guide bracket to thumbscrew #1, thumbscrew #2,
spring, knob, and the body of the needle guide bracket (see Figure
22).
3)
Immerse all parts of the needle guide bracket into the diluted
detergent solution.
4)
Using a syringe flush the cavity and the spiral spring of the needle
guide bracket 5 times under the surface of the detergent solution
with 50 ml diluted detergent. Brush the whole length of the cavity
and the spiral spring of the needle guide bracket at least 5 times
by using an applicable brush.
5)
Wipe the parts of the needle guide bracket under the surface of the
detergent solution with a single-use fluff free soft cloth to remove
all visible soil. Be sure that all grooves of the needle guide bracket
are implemented during the cleaning process.
Содержание C22P
Страница 48: ... 42 Q1E EP1457 Unit mm Fig 23 Dimensions ...
Страница 49: ... 43 Q1E EP1457 Unit mm Fig 24 External view with the attachment ...
Страница 50: ... 44 Q1E EP1457 Unit mm Fig 25 Dimensions with the Needle Guide Bracket ...
Страница 51: ... 45 E Q1E EP1457 Fig 26 External view with Ultra Pro II Needle Guide cover kit Unit mm ...