Console Replacement Instructions
© 2006
Hirschmann ECS
031-300-530-032 Revision F 03/28/06 0032_F53
Page 9
Central Unit Preparation (Not required when replacing the 050-350-061-189 console).
Do not remove
existing cable between console and central unit for installation of the
050-350-061-355
console.
1. Remove existing console cable and strain relief from the central unit and install PG13.5 hole plug
using the existing nut. Remove existing PG11 hole plug. Install new PG11 strain relief using the
existing nut. (Fig. 13)
Cable Assembly Installation (24 350 06 2766 only)
1. Attach inner shield to grounding clamp on the
connection board per diagram. (Fig. 14 and 14a)
2. Install 1/8” and ¼" female terminals on wire
ends and attach to connection board and diode
assy per electrical schematic. (Fig. 15)
3. Reinstall central unit lid and any grounding wires
that were removed.
Grounding of Inner Shield (24 350 06 2766 only)
Inner shield is grounded on connection board.
Cut inner shield back to approximately 1.2 inch. Then fold inner shield back to inner jacket, so the inner
shield is 0.6 inches in length.
Fig. 13
1
Fig. 14
Fig. 14a