55
6
Specifications
6.1 General Specifications
Operating
environment
Indoors, Pollution degree 2, altitude up to 2000 m (6562 ft.)
Operating
temperature
and humidity
Temperature: 0
°
C to 40
°
C (32ºF to 104ºF)
Humidity: 80% RH or less (no condensation)
Storage
temperature
and humidity
Temperature: -10
°
C to 50
°
C (14ºF to 122ºF)
Humidity: 80% RH or less (no condensation)
Standards
Safety: EN61010
EMC: EN61326 Class A
Dielectric
strength
1500 V AC for 1 minute (Sensitivity current: 10 mA), between a batch of L and N
power terminals and the protective ground
5200 V AC for 1 minute (Sensitivity current: 10 mA), between a batch of the
positive and negative terminals and the interface
3270 V AC for 1 minute (Sensitivity current: 10 mA), between a batch of the
positive and negative terminals and the protective ground
Power supply
Rated power
voltage
Commercial power supply 100 V to 240 V AC
(Voltage fluctuations of up to ±10% centered on the rated
supply voltage are allowed.)
(Anticipated transient overvoltage: 2500 V)
Rated power
frequency
50 Hz/60 Hz
Maximum rated
power
150 VA
Display
LED (ERROR, OUTPUT, ALER
T×12, RUN_C×12, RUN
_ positive
×12)
Switch
Power switch, IP address assigning switch
Interface
LAN
Dimensions
Approx.
430W × 132H × 483D
mm (16.93″W× 5.20″D × 19.02″H)
(excluding
protrusions)
Mass
Approx. 10.3 kg (363.3 oz.)
Product
warranty period 1 year
Accessories
Reference: “Accessories” (p.1)
HIOKI SS7081C961-00
Содержание SS7081-50
Страница 7: ...4 HIOKI SS7081C961 00...
Страница 13: ...10 HIOKI SS7081C961 00...
Страница 21: ...18 HIOKI SS7081C961 00...
Страница 57: ...54 HIOKI SS7081C961 00...
Страница 65: ...62 HIOKI SS7081C961 00...
Страница 67: ...64 HIOKI SS7081C961 00...
Страница 69: ...66 Appx 2 Dimensions Diagram 480 460 430 483 15 57 1 Unit mm HIOKI SS7081C961 00...