Query the saved current data 

Syntax

 

Query 

:DATA: CURRent? 

<1 to 12 (NR1)> [,<1 to 15000 (NR1)>]

 

 

Response 

<Measured value (NR3)>, <Measured value (NR3)>••••<Measured value 

(NR3)> 

Description

 

Specifies the channel and the number of data, and reads the current values that have 

been saved. 

All data points are returned if the number of data is not specified. 

If any of the following occurs, an execution error occurs. 

•  While saving 

•  If the number of read data points is greater than the number of data points saved 

•  If no saved data is found 

Do not issue commands or queries while data is being transferred. To cancel a data 

transfer, execute 

*

CLS

Example

 

:DATA: CURR? 1,10 

+2.00001E-04,+1.99999E-04•••+1.99998E-04 

 

(10) 

Self-diagnosis

 

Query the internal temperature

 

Syntax

 

Query 

:SYSTem: TEMPerature? 

<1 to 12 (NR1)/CPU> 

 

Response 

<Measured value>

 

Description

 

Returns the internal temperature for the channel specified using the parameters.

   

When CPU is specified, the temperature of the control board is returned. 

Example

 

:SYST: TEMP? 1 
+4.01 

 

Set and query the overcurrent detection threshold value 

Syntax

 

Command 

[:SOURce]:VOLTage: ILIMit[:LEVel]   

<0.10000 to 1.00000(NR2)/OFF> 

 

Query 

[:SOURce]:VOLTage: ILIMit[:LEVel]? 

 

Response 

<0.10000 to 1.00000(NR2)/OFF>

 

Description 

Sets or queries the overcurrent detection threshold value. 

When the overcurrent detection threshold value is set to OFF, the threshold value 

judgment using the overcurrent is not performed. 

Example

 

:VOLT: ILIM 0.5

 

Sets the overcurrent detection threshold value to 0.5 A.

 

:VOLT: ILIM? 

0.50000

 

The overcurrent detection threshold value is set to 0.5 A. 

 

Set and query the temperature error threshold value 

Syntax

 

Command 

[:SOURce]:VOLTage: TLIMit[:LEVel]

   

<30 to 80 (NR1)> ,<AMP/CPU>

 

 

Query 

[:SOURce]:VOLTage: TLIMit[:LEVel]? 

<AMP/CPU>

 

 

Response 

<30 to 80 (NR1)>

 

Description

 

Sets or queries the threshold value that gives a warning of the temperature error. 

When CPU is set for the last parameter, the temperature error threshold value of the 

control board is set. When AMP is specified, the temperature error threshold value of the 

output board is set. 

Example

 

:VOLT: TLIM 45,AMP 

Sets 45°C for the temperature error threshold value of the 

output board.

 

:VOLT: TLIM? AMP 

45 

The threshold value of the output board that gives a warning of the temperature error has 

been set to 45°C. 

 

 

 

HIOKI SS7081C961-00 

 

Содержание SS7081-50

Страница 1: ...e using the instrument p 2 When using the instrument for the first time Troubleshooting Name and Function of Each Part p 6 Maintenance and Service p 61 Preparation Process p 11 Error Display p 61 HIOK...

Страница 2: ...f 17 3 Generation and Measurement 19 3 1 Voltage Outputs 19 Output constant voltage 19 Changing the output voltage at high speed memory output function 20 3 2 Switching Output Terminals 21 Output term...

Страница 3: ...put for all channels and acquire the measured values 52 Set the output for channel 1 and acquire the measured values 52 Simulate a wire break on channel 2 and then simulate a short circuit on all chan...

Страница 4: ...t ranges express maximum display values Percent of full scale f s means a percentage relative to the maximum display value Reading Displayed value Indicates the value the instrument is presently displ...

Страница 5: ...usness and the hazard levels are classified as follows DANGER Indicates an imminently hazardous situation that will result in death or serious injury to the operator WARNING Indicates a potentially ha...

Страница 6: ...the instrument DANGER Touching any of the high voltage points inside the instrument is very dangerous Do not attempt to modify disassemble or try to repair the instrument Doing so may cause a fire el...

Страница 7: ...4 HIOKI SS7081C961 00...

Страница 8: ...creasingly accurate BMS ICs High accuracy voltage measurement The instrument measures the output voltage using an internal voltmeter with a resolution of 10 V to an accuracy of 0 01 of reading Minuscu...

Страница 9: ...el status display RUN Terminal turns green when a voltage is detected For details see 1 4 Lamp Display p 8 5 Output terminals C terminal Connects the voltage source in series Same potential as the pos...

Страница 10: ...LAN connection The uppermost 24 bits network address are fixed at 192 168 1 Reference 2 4Setting the IP Address p 15 4 Serial number The 9 digit serial number indicates the year first two digits and t...

Страница 11: ...Overcurrent error Output voltage error Abnormal temperature inside the case output board Overrange RUN lamp Output Terminal status C terminal Positive terminal ON NORMAL Lit Lit HIMPEDANCE Lit Unlit...

Страница 12: ...s an Ethernet 100BASE TX If using a 10BASE T or 100BASE TX compatible LAN cable 30 m max to connect to a communications command port using TCP the instrument can be controlled using communications com...

Страница 13: ...10 HIOKI SS7081C961 00...

Страница 14: ...damage the instrument and present an electrical hazard To prevent an electric shock and to maintain the safety specifications of the instrument connect the power cord provided only to an outlet Connec...

Страница 15: ...een the set voltage and the voltage detected across the circuit under test to about 1 with a load current of 1 A while outputting 5 V it is recommended to use a cable length of 2 m or less Example cal...

Страница 16: ...negative terminal or the CHAIN O terminal of the channel in use with highest number may be connected directly to the measuring object 12 channel connection example When connecting CH 1 to the cell wit...

Страница 17: ...erminal of the instrument on the high potential side When connecting the CHAIN terminal to the lowest potential terminal 1 Connect the negative terminal for the channel with the highest number i e the...

Страница 18: ...change the IP address while the instrument is powered on cycle the power Make sure never to turn off 0 or on 255 all the IP address switches Make sure not to connect instruments with the same IP addr...

Страница 19: ...t malfunction or failure Before connecting or disconnecting any measurement cable or LAN cable always turn off the instrument and the computer Failure to do so could result in equipment malfunction or...

Страница 20: ...ming up time 30 minutes is exceeded the blinking instrument OUTPUT display will change to lit IMPORTANT When the instrument is turned on the power supply frequency for the instrument will be set to th...

Страница 21: ...18 HIOKI SS7081C961 00...

Страница 22: ...sible Continuous outputs possible Output time 200 ms 1 A 210 mA 0 210 mA 1 A Reference Communication command Status and event control commands CLS p 41 Output constant voltage Commands may be used to...

Страница 23: ...inal memory output is held Example of 4 point memory output Time T1 voltage V1 Time T2 voltage V2 Time T3 voltage V3 Time T4 voltage V4 Setting memory output 1 Set the memory output value SOURce VOLTa...

Страница 24: ...1 Set the terminal status to ON a batch of all channels OUTPut STATe ON 2 Perform the setting when the terminal is set to ON a batch of all channels or an arbitrary channel OUTPut ON MODE NORMal HIMPe...

Страница 25: ...Terminal status Internal circuit OUTPut STATe OUTPut OFF MODE OFF default value White HIMPEDANCE C terminal The C terminal will be shorted with the negative terminal Positive terminal The positive ter...

Страница 26: ...he ON OFF settings Terminal status OUTPut STATe Internal circuit ON default value RY terminal Connects the terminal to I terminal and O terminal OFF RY terminal Disconnects the terminal from the instr...

Страница 27: ...100 A range The allowable resistance value for the 100 A range is 50 k or greater When the connected resistance value is low the current measured value will go overrange and output will stop referenc...

Страница 28: ...leted If the measurement range is switched If the output terminal is switched If the integral time setting is changed If the output voltage is changed PLC is an abbreviation of Power Line Cycle 1 PLC...

Страница 29: ...st is performed TST In the following cases logging stops automatically If the measurement conditions are changed Execute logging again after finalizing the setting conditions Once the maximum time 12...

Страница 30: ...tput voltage again and turn on output Reference 3 1 Voltage Outputs p 19 3 2 Switching Output Terminals p 21 Output voltage error Output voltage errors occur when the difference between the set output...

Страница 31: ...instrument within the operating temperature and humidity range Has the instrument been installed in an appropriate manner Leave at least 50 mm of space to the left and right of the instrument in fron...

Страница 32: ...ommand messages or query messages 1 Command messages Instructions to control the instrument such as to change or reset the instrument settings E g Command to set the current range SENSe CURRent DC RAN...

Страница 33: ...ers within square brackets may be omitted Either form is valid SENSe CURRent DC RANGe UPPer SENSe CURRent DC RANGe UPPer CURRent RANGe Headers for standard commands This header type begins with an ast...

Страница 34: ...23 34 NR2 Fixed point data e g 1 23 23 45 3 456 NR3 Floating point exponential representation data e g 1 0E 2 2 3E 4 The term NRf format includes all three of the above numeric decimal formats The in...

Страница 35: ...ard Event Register information Output Queue data information bit7 bit6 bit5 bit4 bit3 bit2 bit1 bit0 Unused MSS ESB MAV ESB0 Unused Unused Unused Status Byte Register STB Logical sum bit7 bit6 bit5 bi...

Страница 36: ...output of EV and EN Status Byte Register AND output of EV and EN Logical sum Logical sum Output data Serial poll Standard Event Status Register AND output of EV and EN Returns the status when STB is...

Страница 37: ...nd contains a syntactic or semantic error Program header error Incorrect number of data parameters Invalid parameter format Received a command not supported by the instrument Bit 4 EXE Execution error...

Страница 38: ...is a 16 bit register When any bit in one of these Event Status Registers enabled by its corresponding Enable Register is set to 1 the following happens For Status Query Register Bit 3 ESB0 of the Stat...

Страница 39: ...is incorrect Query error If the controller cannot receive or if the instrument cannot send response messages Execution error If other than the designated character data or numerical data has been set...

Страница 40: ...o 12 Sets the output terminal during output on OUTPut ON MODE Channel number 1 to 12 NORMAL HIMPEDANCE ZERO Queries the output terminal during output on OUTPut OFF MODE HIMPedance ZERO Sets the output...

Страница 41: ...0000 OFF Sets the overcurrent detection threshold value SOURce VOLTage ILIMit LEVel 0 10000 to 1 00000 OFF Queries the overcurrent detection threshold value SOURce VOLTage TLIMit LEVel 30 to 80 AMP CP...

Страница 42: ...00 to 5 0250E 00 Description Sets or queries the output voltage for the channel specified using the last parameter Command If channel numbers are omitted the same settings are made in all the channels...

Страница 43: ...hold value 2 0 mV Threshold value of temperature error inside the case control board 50 C Threshold value of temperature error inside the case output board 70 C Interval for which error detection is d...

Страница 44: ...mand CLS Description Clears the status byte Note Implementing this command while data is being transferred using the logging acquisition query will stop the data transfer The saved data is not cleared...

Страница 45: ...is rounded to the nearest integer The value of the unused bit expressed by such as bit 6 is ignored When the power is turned on such bit is initialized to 0 Query Returns the contents of the SRER set...

Страница 46: ...uery is executed or the power is turned off then on again Executing this command after an overcurrent has been detected will cancel the output stopped state Set and query the Enable Register of the St...

Страница 47: ...ror has occurred in channel 1 Note When bit 5 VOLT_ERR of the Status Query Register is set to 1 the bit for the channel that causes the error is also set to 1 This bit is not cleared unless CLS RST or...

Страница 48: ...en the output terminal is set to ON the set voltage is output to the C and positive terminals for all 12 channels OUTP ON MODE 1 NORMAL The setting is made so that the set voltage is output to the C a...

Страница 49: ...an arbitrary output voltage for all channels at the same time Command If channel numbers are omitted the same settings are made in all the channels The minimum resolution of the voltage setting is 0 0...

Страница 50: ...V in 3 0 s interval Memory 4 Output for 0 0 V in 1 0 s interval VOLT MEM TABL 1 0 500 0 00000E 00 2 000 4 20000E 00 3 000 2 00000E 00 1 000 0 00000 E 00 Returns the memory output value for channel 1 S...

Страница 51: ...mmas are returned The minimum effective resolution of the voltage data is 0 00001 V Example FETC VOLT 1 1 50000E 00 Query the output current value Syntax Query FETCh CURRent 1 to 12 NR1 Response Measu...

Страница 52: ...ified time has elapsed from the start of saving When the saving time is omitted the current and voltage values are saved for 12 hours continuously until the operation stops In addition an execution er...

Страница 53: ...on threshold value Syntax Command SOURce VOLTage ILIMit LEVel 0 10000 to 1 00000 NR2 OFF Query SOURce VOLTage ILIMit LEVel Response 0 10000 to 1 00000 NR2 OFF Description Sets or queries the overcurre...

Страница 54: ...Sets or queries the error detection disabled time when switching from the 1 A range to the 100 A range Example VOLT LIM DEL 1 Sets the error detection time to 1 second VOLT LIM DEL 1 000 The error de...

Страница 55: ...1 Sets channel 1 to the 1 A range VOLT 3 3 1 Outputs 3 3 V to channel 1 FETC VOLT 1 3 30003E 00 Acquires the measured voltage value for channel 1 FETC CURR 1 5 20000E 03 Acquires the measured current...

Страница 56: ...ll channels to 3 3 V DATA STAT 1 Start logging DATA STAT 0 DATA POIN 1 4 DATA CURR 1 2 00000E 10 2 00000E 10 2 00000E 10 3 00000E 10 DATA VOLT 1 3 30003E 00 3 30003E 00 3 30003E 00 3 30003E 00 Stop lo...

Страница 57: ...54 HIOKI SS7081C961 00...

Страница 58: ...etween a batch of the positive and negative terminals and the interface 3270 V AC for 1 minute Sensitivity current 10 mA between a batch of the positive and negative terminals and the protective groun...

Страница 59: ...0 ms Time to next output reference value 5 s when outputting 1 A at 5 V for 200 ms Output settling time Time until output reaches 99 9 of final value 10 ms reference value 1 F connection 100 A range O...

Страница 60: ...ditional error temperature coefficient From 0 C to 18 C and from 28 C to 40 C add following value 0 05 output accuracy per C Output resistance 3 m or less Does not include terminal s contact resistanc...

Страница 61: ...e measured current value Settings Logging measurement ON OFF Automatic stop time 1 00 s to 99 99 s Default setting Logging measurement OFF Memory output function Operation Output will vary automatical...

Страница 62: ...trigger the no output state No voltage output will be generated while in the no output state Setting Threshold 0 1 A to 1 0 A Default value 1 0 A Output voltage error detection Operation If the differ...

Страница 63: ...Connector RJ 45 Communications Configuration equipment status acquisition settings acquisition via communication commands Settings IP address 192 168 1 xxx only the xxx portion can be changed Subnet...

Страница 64: ...en a soft cloth moistened with water or a neutral detergent and wipe the instrument clean Wipe the display gently with a soft dry cloth Clean the vents periodically to avoid blockage 7 1 Error Display...

Страница 65: ...62 HIOKI SS7081C961 00...

Страница 66: ...distribution 3 The name of the author may not be used to endorse or promote products derived from this software without specific prior written permission THIS SOFTWARE IS PROVIDED BY THE AUTHOR AS IS...

Страница 67: ...64 HIOKI SS7081C961 00...

Страница 68: ...truments in series enabled 13 or more channels to be connected 1000 V max As the output voltage and output current for each channel are measured simultaneously no measured values are unexpectedly lost...

Страница 69: ...66 Appx 2 Dimensions Diagram 480 460 430 483 15 57 1 Unit mm HIOKI SS7081C961 00...

Страница 70: ...er than 6 mm When removing the rack mounting tool and restoring the original configuration use the same screws as were used for the original mounting M3 6 mm If installing the instrument on a separate...

Страница 71: ...68 HIOKI SS7081C961 00 Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...

Отзывы: