![Hioki SM7860 Series Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/hioki/sm7860-series/sm7860-series_instruction-manual_2137403053.webp)
49
Basic Specifications
Accuracy Specifications
Conditions of guaranteed accuracy
Guaranteed accuracy period
1 year
Guaranteed accuracy period
after adjustment made by Hioki
1 year
Temperature and humidity for
guaranteed accuracy
23°C±5°C (73°F±9°F), 80% RH or less
Warm-up time
at least an hour
Power supply frequency
50 Hz/60 Hz ±2 Hz
Voltage generation accuracy
Output voltage accuracy
±2% of setting ±0.5 V (no load)
Error between channels
±0.01 V or less (no load, between outputs from same circuit)
Voltage monitor accuracy
±2% of output voltage ±0.5 V
Current limit accuracy
±10% of setting ±1 mA
Limitations
Voltage application target
Multi-layer ceramic capacitor
Number of charging channels
Up to 8 channels/circuit
Operating conditions
Charging interval must be greater than or equal to the time
calculated from the graphs shown in Figure 1 and Figure 2 below.
(Continuous charging is not supported.)
0.1
1.0
10.0
100.0
10
100
1000
6.3 V 10 V 16 V 25 V
50 V 100 V
4 V
Applied voltage
Charging interval [ms]
Capacitance [µF]
Applied voltage 500 V
Applied voltage 250 V
Figure 1: Charging Interval by Applied Voltage and Capacitance
(SM7860-51, -53, -55, -57, -58, -61, -63, -65, -67, -68)