26
Fig. 7.1.1
Fig. 7.1.1 shows the pin arrangement label for the DC OUT (OPTION)
connector on the rear of the unit.
A DC measuring instrument whose input impedance is greater than
1M
Ω
like a HIOKI chart recorder can be connected to the connector.
For permanent data recording, a HIOKI chart recorder is recommended
for better technical follow-up, including a supply of a variety of
recording charts and technical service.
3) Output voltage
These tables show the relations between displayed value of SM-8216
super megohmmeter and output voltage of RP-8000.
Target voltage 10V
Displayed value
0.05 0.25 0.5
Output voltage
1V 5V 10V
Target voltage 25V
Displayed value
0.125 0.625 1.25
Output voltage
1V 5V 10V
Target voltage 50V
Displayed value
0.25 1.25 2.5
Output voltage
1V 5V 10V
Target voltage 100V
Displayed value
0.5 2.5 5
Output voltage
1V 5V 10V
Target voltage 250V
Displayed value
1.25 6.25 12.5
Output voltage
1V 5V 10V