4
Fuse Replacement
The fuse is inserted in the FUSE holder (Fig. 1.1.2) on the rear of the unit.
Remove the cap, and replace the fuse with a new one with a correct
amperage.
To remove the cap of the fuse holder, use a 4 mm Phillips screwdriver, and
turn the cap counterclockwise.
To set the cap in position, insert the cap, holding the fuse into the holder,
and turn it with the screwdriver.
Fig. 1.1.2 Fuse Holder and Fuse Replacement
1.5 Grounding the Chassis
To avoid an electrical shock accident, connect the ground prong of
the power
cord to the ground post of the AC line system.
The round prong of the 3-prong plug of the power cord is the ground prong.
It is recommended to use an AC line socket with its third contact grounded
for connection of the accessory 3-prong power cord.
To use the accessory 3-prong to 2-prong adapter to connect the power cord
to a 2-contact AC line socket, make sure to ground the green tab of the
adapter.
WARNING
To replace or check the fuse, make sure to disconnect the power cord
from the AC line socket. If not, there is a fear of electrical shock.
WARNING
To prevent an accident, connect the ground prong of the power cord
Plug to the ground post of the AC line system. If it is impossible to
ground the ground prong of the power cord plug, be sure to connect
the GND terminal on the rear of the unit.