Fig. 1.1.1 LINE SELECTOR Switch Setting
The super megohmmeter can be operated from one of the following AC
power source when the LINE SELECTOR switches are set accordingly.
LINE SELCTOR Setting Acceptable Line Voltage Range
100 V 90 V to 110V (100 V
±
10%)
120 V 108 V to 132 V (120 V
±
10%)
220 V 198 V to 242 V (220 V
±
10%)
240 V 216 V to 250 V (240 V +10V,
‑
10%)
1.4 Checking the Fuse Amperage
The fuse holder of the instrument contains a time lag fuse of the following
amperage:
CAUTION
When the LINE SELECTOR switch position is changed, fuse
amperage must be changed to meet the requirements of the new AC
line voltage.
Incorrect fuse amperage will cause a failure of the instrument.
Time Lag Fuse
AC Line Voltage Fuse Amperage
100 V/120 V 0.4 A
220 V/240 V 0.2 A
100V
120V
220V
240V
LINE SELECTOR