17
DANGER
When the MEASURE/CHARGE/DISCHARGE function selector is
to the CHARGE or MEASURE position, never touch the Rx ‘-‘/’+’
measuring terminals nor the sample in connection. This is
because the voltage set with the VOLTAGE selector is directly
applied to the above-mentioned sites.
An inadvertent touch of any of the mentioned sites will give an
electrical shock to the operator.
8) Setting the Measuring Voltage
(21) Set the VOLTAGE selector switch to the position in accordance with
your measurement program.
(22) If there is no idea about the measuring voltage to use, set the
VOLTAGE selector to the lowest voltage available (10 V) to try the first
measurement, then, increase the voltage in response with the result of
the measurement.
9) Setting the Measuring Range
(23) Set the RANGE selector to the position in response to the set position
of the VOLTAGE selector.
(24) If there is no idea about the insulation resistance range to use, set
the RANGE selector to the lowest range available (
×
1) to try the first
measurement, then, widen the range in response with the result of
the measurement.