7.1 FT3470-50 Magnetic Field Tester
7.1.3 Output Function
Supporting function
Magnetic flux density (T), Exposure level (%)
Output type
MON : 3-axis waveform output
REC : Resultant RMS level output (output via the X-axis.)
Exposure level output (output via the X-axis.)
Output rate
0.1 mV/display value count
An output rate based on the magnetic flux density unit T is used, even if the
magnetic flux density unit is G or A/m.
Output accuracy
Measurement function/
Range
Output rate
Waveform output (MON)
Level value (REC)
Magnetic flux
density
r
0
AC 100 mV/
P
T
DC 100 mV/
P
T
r
1
AC 10 mV/
P
T
DC 10 mV/
P
T
r
2
AC 1 mV/
P
T
DC 1 mV/
P
T
r
3
AC 100 mV/mT
DC 100 mV/mT
Exposure level
r
0
-
DC 10 mV/%
r
1
-
DC 1 mV/%
Output accuracy (MON) is defined under the condition that the input is sine wave with no dis-
tortion.
Type
Measurement
function
Measurement
Mode
Output accuracy
Effective fre-
quency range
REC
Magnetic flux
density
10 Hz-400 kHz
±3.5%rdg.±3 mV
50 Hz to 100 kHz
10 Hz-2 kHz
50 Hz to 1 kHz
2 kHz-400 kHz
5 kHz to 100 kHz
Exposure level
General Public/
Occupational
±3.5%rdg.±3 mV
50 Hz to 1 kHz
±5.0%rdg.±3 mV
1 kHz to 100 kHz
MON
Magnetic flux
density
10 Hz-400 kHz
±3.5%rdg.±10 mV
50 Hz to 100 kHz
10 Hz-2 kHz
50 Hz to 1 kHz
2 kHz-400 kHz
5 kHz to 100 kHz