7.1 FT3470-50 Magnetic Field Tester
103
2
3
4
5
6
7
8
9
10
11
12
13
7
3. Exposure level (General Public, Occup.)
Accuracy
specifications
Smoothed edges
Exposure level is an IEC/EN62233-defined measurement technique apply-
ing time domain evaluation to magnetic flux density as described by the
ICNIRP 2010 guidelines.
It differs by a maximum of 3 dB from the ICNIRP 2010 corner frequency.
4. Characteristic
Crest factor
3 or less (But exposure level (occupational) for r1 and r2 is 1.45 or less.)
Effect of expo-
sure
to power-fre-
quency
fields
In a 20 kV/m, 60 Hz field at resultant RMS value R
Effect of tester
on magnetic field
With magnetic sensor directly connected to the tester
Noise level
(reference value)
RMS integral time
(reference value)
Approx. 120 ms (slow function off)
Approx. 1 s (slow function on)
Measured axes
Range
Effective
measuring range
Accuracy
X, Y, Z
r
0
0.50 to 20.00%
Smoothed edges 50 Hz to 1 kHz
±3.5%rdg.±0.5%f.s.
Smoothed edges 1 kHz to 100 kHz
±5.0%rdg.±0.5%f.s.
r
1
05.0 to 200.0%
R
r
0
1.00 to 34.64%
r
1
10.0 to 346.4%
Magnetic field sensor
Influence of resultant RMS value R
Magnetic flux density
㸦
10 Hz to 400 kHz
㸧
Exposure level
(General Public)
100 cm
2
Magnetic Field Sensor
0.02
P
T or less
0.1% or less
3 cm
2
Magnetic Field Sensor
0.15
P
T or less
0.6% or less
Magnetic field sensor
Influence of resultant RMS value R
100 cm
2
Magnetic Field Sensor
±1.5%rdg. or less
3 cm
2
Magnetic Field Sensor
±2.5%rdg. or less
Magnetic field sensor
Measurement function/mode
Resultant RMS value R
100 cm
2
Magnetic Field Sensor
Magnetic flux density
(10 Hz to 400 kHz)
0.050
P
T
3 cm
2
Magnetic Field Sensor
0.150
P
T
www.
.com
1.800.561.8187