60
General Specifications
Dust resistance and
water resistance
IP20 (When measuring voltage under completely dry conditions, when measuring
current flowing through a hazardous live conductor under completely dry
conditions)
IP50 (When measuring resistance under completely dry conditions)
IP54 (When measuring current flowing through an insulated conductor, during
storage)
The protection rating for the enclosure of this instrument (based on EN60529) is
IP20*, IP50*, or IP54*.
* IP20, IP50, IP54:
This indicates the degree of protection provided by the enclosure of the device for
use in hazardous locations, entry of solid foreign objects, and the ingress of water.
2
: Protected against access to hazardous parts with a fingers. The equipment
inside the enclosure is protected against entry by solid foreign objects larger
than 12.5 mm in diameter.
5
: Protected against access to hazardous parts with wire measuring 1.0 mm
in diameter. Dustproof type (The penetration of dust cannot be prevented
completely, but quantities of dust that may hinder the stated operation of
equipment or safety cannot penetrate the enclosure.)
0
: The equipment inside the enclosure is not protected against the harmful effects
of water.
4
: The equipment inside the enclosure is protected against the harmful effects of
water splashed against the enclosure from any direction.
Standards
Safety
EN 61010
EMC
EN 61326
HIOKI CM4371C961-00
Содержание CM4371-50
Страница 2: ...HIOKI CM4371C961 00...
Страница 6: ...iv Contents HIOKI CM4371C961 00...
Страница 112: ...106 Accuracy table HIOKI CM4371C961 00...
Страница 120: ...Index 114 HIOKI CM4371C961 00...
Страница 121: ...HIOKI CM4371C961 00...
Страница 122: ...HIOKI CM4371C961 00...
Страница 123: ...HIOKI CM4371C961 00...
Страница 124: ...www hioki com HIOKI CM4371C961 00...