109
Error and Operation Displays
4.2 Error and Operation Displays
Error display
Description
Remedy
v.up
The instrument’s firmware is
being upgraded.
Do not remove the batteries until the upgrade
completes.
Err 001
ROM error (program)
When the error appears on the display, it is
necessary to repair the instrument. Please
contact your authorized Hioki distributor or
reseller.
Err 002
ROM error (adjustment data)
Err 004
Memory error
Err 005
ADC error
(hardware malfunction)
Err 008
Z3210 communication error
(the Z3210 is malfunctioning or
is not properly connected)
Perform the following steps. (p. 23)
Disconnect and then reconnect the Z3210.
If you have another Z3210, replace the unit in
question with that one.
If the error continues to be displayed, the
instrument needs to be repaired.
Please contact your authorized Hioki
distributor or reseller.