30
────────────────────────────────────────────────────
5.3 Standard Commands
────────────────────────────────────────────────────
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
PON
URQ
CME
EXE
DDE
QYE
RQC
OPC
Standard event status enable register (SESER)
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
PON
URQ
CME
EXE
DDE
QYE
RQC
OPC
Standard event status register (SESR)
*
ESE?
■
Queries the standard event status enable register.
*
ESR?
■
Queries the contents of the standard event status register (SESR).
*
IDN?
■
Queries manufacturer’s name, model name, and software version.
Syntax
*
ESE?
Function
Returns the setting contents of SESER as a numerical value in NR1 format
between 0 and 255.
Example
Response
If headers are on
*ESE 20
If headers are off
20
Bits 2 and 4 of SESER have been set to 1.
Error
If the response message is longer than 300 bytes, a query error is generated.
Syntax
*
ESR?
Function
・
Returns the contents of the standard event status register (SESR) as a
numerical value in NR1 format between 0 and 255, and then clears standard
event status register.
・
No header is affixed to the response message.
Example
Response
32
Bit 5 of SESR has been set to 1.
Error
If the response message is longer than 300 bytes, a query error is generated.
Syntax
*
IDN?
Function
・
The response consists of the name of the manufacturer of the unit, the model
name, and the software version.
・
No header is affixed to the response message.
First field Manufacturer’s name
Second field Model name
Third field Serial number (Not used - always zero)
Fourth field Software version
Example
Response
HIOKI,3157,0,V01.01
Error
If the response message is longer than 300 bytes, a query error is generated.