29
────────────────────────────────────────────────────
5.3 Standard Commands
────────────────────────────────────────────────────
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
PON
URQ
CME
EXE
DDE
QYE
RQC
OPC
Standard event status enable register (SESER)
5.3 Standard Commands
*
CLS
■
Clears the status byte register and the event registers.
*
ESE
■
Sets the standard event status enable register.
Syntax
*
CLS
Function
・
Clears all the event registers (SESR, ESR0) associated with the bits of the
status byte register. Accordingly, also clears the status byte register.
・
This has no effect upon the output queue, the various enable registers, or bit 4
(the MAV bit) of the status byte register.
Error
If the data parameters are set after this command, a command error occurs.
Syntax
*
ESE
_<data>
<data>
Numerical data in NR1 format between 0 and 255
Function
・
Sets the standard event status enable register (SESER) to a bit pattern which
is used to mask the standard event status register (SESR).
・
The numerical value can be in NRf format, but any digits after the decimal
point will be rounded.
・
When the power is turned on, the data is reinitialized to zero.
Example
Transmission
*
ESE
20
Bits 2 and 4 of SESER are set to 1.
Error
If <data> is other than numerical value described above, an execution error
occurs.