background image

5.3  Timing Chart

76

Description

Time

t1

ERR Output response 
time

*1

1.5 ms

t2

Measurement trigger 
pulse width

0.5 ms min.

t3

Delay Time

per setting

™

See Section 4.2.2 Trigger Delay Settings (Page 52).

t4

Measurement time

*2

Ω

V mode

EX.FAST

6.8 ms

FAST

22.8 ms

MEDIUM

82.8 ms
68.8 ms

SLOW

257.8 ms
251.2 ms

Ω

 mode or V mode

EX.FAST

3.4 ms

FAST

11.4 ms

MEDIUM

41.4 ms (50 Hz line 
frequency setting)
34.4 ms (60 Hz line 
frequency setting)

SLOW

156.4 ms (50 Hz line 
frequency setting)
149.8 ms (60 Hz line 
frequency setting)

t5

Calculation time

*3

0.3 ms

t6

EOC Output pulse width

When the external trigger is selected
HOLD setting

: Holds until the next trigger is detected

PULSE setting : Remains only for the specified pulse width

™

See Section 5.2.5 Instrument Settings (Page 74).

When the internal trigger is selected
HOLD setting

: EX.FAST 1 ms, FAST 5 ms, MEDIUM 20 ms, SLOW 50 ms

PULSE setting : Remains only for the specified pulse width

*1: For details, see “5.2.4 ERR Output (Page 73).”

*2

About t4 measurement time

When averaging is enabled, the running average is obtained with internal triggering, so
measurement time t4 does not change. The measurement time for external triggering is as
follows:
With SLOW sampling

Ω

V

(t4 - 57.8) X n + 57.8 ms (50 Hz)
(t4 - 51.2) X n + 51.2 ms (60 Hz)

Ω

 or V

(t4 - 56.4) X n + 56.4 ms (50 Hz)
(t4 - 49.8) X n + 49.8 ms (60 Hz)

With other than SLOW sampling

Ω

V

(t4 - 2.8) X n + 2.8 ms

Ω

 or V

(t4 - 1.4) X n + 1.4 ms

(n represents the number of values averaged)

*3

About t5 calculation time

In the following cases, add the indicated times to calculation time t5:

When the Statistical Calculation function is enabled

0.3 ms

When the reference value/tolerance method of 
comparator decision is selected

0.15 ms

Содержание 3561

Страница 1: ...BATTERY HiTESTER INSTRUCTION MANUAL 3561 3561 01 ...

Страница 2: ......

Страница 3: ......

Страница 4: ... Judgment Results 49 4 1 9 Switching Between Measurement Value and Comparator Setting Displays 50 4 2 Trigger Function 51 4 2 1 Trigger Source Settings 51 4 2 2 Trigger Delay Settings 52 4 3 Averaging Function 53 4 4 Statistical Calculation Functions 54 4 5 Memory Function 58 4 6 Key Lock Function 60 4 7 Panel Save Function 61 4 8 Panel Load Function 62 4 9 Self Calibration 63 4 10 Measurement Val...

Страница 5: ...7 2 2 GP IB Specifications Model 3561 01 only 88 7 3 Selecting the Connections and Protocol 89 7 3 1 Attaching the Connector 89 7 3 2 Selecting the Interface 91 7 4 Communication Methods 92 7 4 1 Message Format 92 7 4 2 Output Queue and Input Buffer 97 7 4 3 Status Byte Register 98 7 4 4 Event Registers 100 7 4 5 Initialization Items 103 7 4 6 Local Function 103 7 5 Message List 104 7 5 1 Standard...

Страница 6: ...4 Appendix_________________________________165 Appendix 1 Precautions for Making Custom Test Leads 165 Appendix 2 AC Four terminal Method 167 Appendix 3 Synchronous Detection System 168 Appendix 4 Configuration and Extension of the Test Leads 169 Appendix 5 Effect of Eddy Currents 170 Appendix 6 Calibration Procedure 171 Appendix 7 Test Lead Options 172 Appendix 8 Rack Mounting 173 Appendix 9 Dime...

Страница 7: ......

Страница 8: ......

Страница 9: ......

Страница 10: ......

Страница 11: ......

Страница 12: ......

Страница 13: ......

Страница 14: ......

Страница 15: ......

Страница 16: ......

Страница 17: ......

Страница 18: ......

Страница 19: ......

Страница 20: ......

Страница 21: ......

Страница 22: ......

Страница 23: ......

Страница 24: ......

Страница 25: ......

Страница 26: ......

Страница 27: ......

Страница 28: ......

Страница 29: ......

Страница 30: ......

Страница 31: ......

Страница 32: ......

Страница 33: ......

Страница 34: ......

Страница 35: ......

Страница 36: ......

Страница 37: ......

Страница 38: ...H and SOURCE L is 50 Ω or more in the 300 mΩ range or 500 Ω or more in the 3 Ω range When the resistance between SENSE H and SENSE L is greater than about 20 Ω However if the capacitance of the test leads is 1 nF or higher the measurement fault may not be detected When a bad contact results from damage excessive wear or impurities on the test leads If the circuit protection fuse is blown See Secti...

Страница 39: ......

Страница 40: ......

Страница 41: ......

Страница 42: ......

Страница 43: ......

Страница 44: ......

Страница 45: ......

Страница 46: ......

Страница 47: ......

Страница 48: ......

Страница 49: ......

Страница 50: ......

Страница 51: ......

Страница 52: ......

Страница 53: ......

Страница 54: ......

Страница 55: ......

Страница 56: ......

Страница 57: ......

Страница 58: ......

Страница 59: ......

Страница 60: ......

Страница 61: ......

Страница 62: ......

Страница 63: ......

Страница 64: ......

Страница 65: ......

Страница 66: ......

Страница 67: ......

Страница 68: ......

Страница 69: ......

Страница 70: ......

Страница 71: ......

Страница 72: ......

Страница 73: ......

Страница 74: ...4 12 Reset Function 68 ...

Страница 75: ......

Страница 76: ......

Страница 77: ......

Страница 78: ... INDEX signal is output during the Trigger Wait Delay Self Calibration and Calculation states This signal is not output while measuring the resistance of test objects This signal transitions from Off to On to indicate that the test object can be removed EOC This signal indicates the end of a measurement End Of Conversion This signal indicates when comparator judgment results and ERR output when SY...

Страница 79: ......

Страница 80: ......

Страница 81: ......

Страница 82: ...Instrument Settings Page 74 When the internal trigger is selected HOLD setting EX FAST 1 ms FAST 5 ms MEDIUM 20 ms SLOW 50 ms PULSE setting Remains only for the specified pulse width 1 For details see 5 2 4 ERR Output Page 73 2 About t4 measurement time When averaging is enabled the running average is obtained with internal triggering so measurement time t4 does not change The measurement time for...

Страница 83: ......

Страница 84: ......

Страница 85: ......

Страница 86: ......

Страница 87: ......

Страница 88: ......

Страница 89: ......

Страница 90: ......

Страница 91: ......

Страница 92: ...9 98mOhm 16 0 2103 Ohm With erroneous measurement values 10 O F O F 11 O F O F 12 13 Invalid Invalid 14 O F 12 0097 V 15 19 82mOhm O F With the Comparator ON 95 105 80mOhm Lo 0 0000 V IN 96 213 15mOhm Hi 97 213 12mOhm IN 98 213 11mOhm Lo 99 10 0072 V Hi 100 10 0071 V IN 101 10 0070 V Lo 102 O F Hi O F Hi 103 3 11mOhm Lo O F Lo 104 With the relative value comparison method reference value and toler...

Страница 93: ......

Страница 94: ......

Страница 95: ......

Страница 96: ......

Страница 97: ......

Страница 98: ......

Страница 99: ......

Страница 100: ...n this way and if one command contains an error all subsequent messages up to the next terminator will be ignored A query error occurs if a query command is combined with an immediately following semicolon and subsequent command 2 Header Separator In a message consisting of both a header and data the header is separated from the data by a space 3 Data Separator In a message containing multiple dat...

Страница 101: ......

Страница 102: ......

Страница 103: ......

Страница 104: ......

Страница 105: ......

Страница 106: ...alid parameter format Received a command not supported by the instrument Bit 4 EXE Execution Error This bit is set to 1 when a received command cannot be executed for some reason The specified data value is outside of the set range The specified setting data cannot be set Execution is prevented by some other operation being performed Bit 3 DDE Device Dependent Error This bit is set to 1 when a com...

Страница 107: ......

Страница 108: ......

Страница 109: ......

Страница 110: ......

Страница 111: ......

Страница 112: ...ltage comparator lower threshold setting 121 CALCulate LIMit VOLTage LOWer Lower threshold Queries the voltage comparator lower threshold setting 121 CALCulate LIMit RESistance REFerence Reference value Sets the resistance comparator reference value 122 CALCulate LIMit RESistance REFerence Reference value Queries the resistance comparator reference value 122 CALCulate LIMit VOLTage REFerence Refer...

Страница 113: ......

Страница 114: ...2 SYSTem BACKup Backups current measurement configuration 132 Header Present SYSTem HEADer 1 0 ON OFF Sets header present 133 SYSTem HEADer ON OFF Queries the header present setting 133 ERR Output SYSTem ERRor SYNChronous ASYNchronous Sets error output timing 133 SYSTem ERRor SYNChronous ASYNchronous Queries the error output timing setting 133 Terminator SYSTem TERMinator 0 1 Sets the terminator 1...

Страница 115: ......

Страница 116: ......

Страница 117: ......

Страница 118: ...shed processing Wait for Pending Commands to Finish Syntax Command WAI Description Command The instrument waits until all prior commands finish before executing any subsequent commands Note The WAI command is supported because it is defined in IEEE 488 2 1987 but because all Model 3561 3561 01 device specific commands are sequential types this command has no actual affect Clear the Status Byte and...

Страница 119: ......

Страница 120: ... zero Example Command SRE 33 Set SRER bits 0 and 5 to 1 Query SRE Response 33 SRER bits 0 and 5 have been set to 1 128 64 32 16 8 4 2 1 bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0 unused 0 ESB MAV unused unused ESE1 ESE0 Query the Status Byte and MSS Bit Syntax Query STB Response 0 to 255 NR1 Description Query The contents of the STB are returned as an NR1 value 0 to 255 The response message h...

Страница 121: ......

Страница 122: ...0 to 3 1 Query RESistance RANGe Response 300 00E 3 3 0000E 0 NR3 Example Command RES RANG 120E 3 Selects the most suitable resistance measurement range for measuring 120 mΩ Query Response RES RANG 300 00E 3 The current resistance measurement range is 300 mΩ Note Changing the resistance measurement range clears stored measurement data memory function Set and Query the Voltage Measurement Range Synt...

Страница 123: ......

Страница 124: ...CALCulate AVERage 2 to 16 Query CALCulate AVERage Response 2 to 16 NR1 Example Command CALC AVER 10 Query CALC AVER Response 10 Set and Query the Comparator Syntax Command CALCulate LIMit STATe 1 0 ON or OFF Query CALCulate LIMit STATe Response ON or OFF Example Command CALC LIM STAT ON Query CALC LIM STAT Response ON Note When the Comparator function is enabled auto ranging is disabled Switching ...

Страница 125: ......

Страница 126: ... threshold is set to 2 8593 Ω Query CALC LIM RES UPP Response 28593 Note The value is sent as a whole integer count To set 120 53 mΩ with the 300 mΩ range send the following CALC LIM RES UPP 12053 Voltage Measurement Syntax Command CALCulate LIMit VOLTage UPPer Upper threshold Query CALCulate LIMit VOLTage UPPer Response Upper threshold Upper threshold 0 to 999999 NR1 Example Command CALC LIM VOLT...

Страница 127: ......

Страница 128: ...shold is set to 0 5076 Ω Query CALC LIM RES REF Response 5076 Note The value is sent as a whole integer count To set 120 53 mΩ with the 300 mΩ range send the following CALC LIM RES REF 12053 Voltage Measurement Syntax Command CALCulate LIMit VOLTage REFerence Reference value Query CALCulate LIMit VOLTage REFerence Response Reference value Reference value 0 to 999999 NR1 Example Command CALC LIM VO...

Страница 129: ......

Страница 130: ...CALC STAT STAT Response ON About the Statistical Calculation function Data samples can be acquired by the following three methods Press the TRIG key Apply an EXT I O TRIG signal Send the TRG command The CALCulates STATistics STATe command does not clear calculation results When the valid data count is zero σn 1 returns 0 Clearing calculation results does not disable the Statistical Calculation fun...

Страница 131: ......

Страница 132: ...imum Response Maximum value NR3 Data No of Maximum value NR1 Example Query CALC STAT VOLT MAX Response 1 3924E 0 1 Query the Minimum value Resistance Measurement Syntax Query CALCulate STATistics RESistance MINimum Response Minimum value NR3 Data No of Minimum value NR1 Example Query CALC STAT RES MIN Response 294 88E 3 8 Voltage Measurement Syntax Query CALCulate STATistics VOLTage MINimum Respon...

Страница 133: ......

Страница 134: ...ES CP Response 0 04 0 04 Voltage Measurement Syntax Query CALCulate STATistics VOLTage CP Response Cp NR2 CpK NR2 Example Query CALC STAT VOLT CP Response 0 91 0 00 Set and Query the Memory Function State Syntax Command MEMory STATe 1 0 ON OFF Query MEMory STATe Response ON OFF Example Command MEM STAT ON Query MEM STAT Response ON Clear Instrument Memory Syntax Command MEMory CLEAr ...

Страница 135: ......

Страница 136: ... 0 ON or OFF Query SYSTem DATAout Response ON or OFF ON Measured values are output automatically when a trigger occurs OFF Measured values are not output Example Command SYST DATA OFF Query SYST DATA Response OFF Note This function is convenient when you want to obtain measured values by applying EXT I O trigger input When this function is enabled and a footswitch is connected to the TRIG terminal...

Страница 137: ......

Страница 138: ... Load Measurement Values Syntax Command SYSTem SAVE 1 to 126 SYSTem LOAD 1 to 126 Note Attempting to load a panel number that has not been saved results in an execution error Up to 126 measurement configurations can be saved and loaded Refer to Panel Save and Load Functions for details Backup Current Measurement Configuration Syntax Command SYSTem BACKup Description Command The current measurement...

Страница 139: ......

Страница 140: ...t9 bit8 bit7 bit6 bit5 bit4 bit3 bit2 bit1 bit0 OUT9 OUT8 OUT7 OUT6 OUT5 OUT4 OUT3 OUT2 OUT1 OUT0 Pin No 36 18 35 17 34 16 33 15 32 14 EXT I O Input Syntax Query IO IN Response 0 to 31 NR1 Description Query Signals at the EXT I O IN0 to IN4 input terminals are read at the leading edge Each bit edge data is cleared upon reading by this query A bit is set when the leading edge short between each sig...

Страница 141: ......

Страница 142: ......

Страница 143: ......

Страница 144: ...r source Query Queries the trigger source selection Example Command TRIG SOUR IMM Sets the trigger source to internal triggering Query Response TRIG SOUR IMMEDIATE The trigger source is set to internal triggering Enable Disable and Query Trigger Delay Syntax Command TRIGger DELay STATe 1 0 ON or OFF Query TRIGger DELay STATe Response ON or OFF ON Trigger delay enabled OFF Trigger delay disabled Ex...

Страница 145: ......

Страница 146: ...nt Example Query READ Response 289 68E 3 1 3921E 0 ΩV mode Measured resistance is 289 68 mΩ and voltage is 1 3921 V Error This command causes an execution error if issued during the Continuous Measurement state after INITIATE CONTINUOUS ON Note The next command does not execute until measurement is finished When the trigger source is external the TRG command does not trigger measurement See Sectio...

Страница 147: ......

Страница 148: ...her as PASS FAIL This instrument judges resistance and voltage independently Also when the Comparator function is enabled ON auto ranging is disabled OFF Voltage Limiter This instrument does not include a voltage limiter function limiting open terminal voltage to 20 mV This instrument s open terminal voltage is 7 V peak dropping to a few millivolts when the test leads are connected to a test objec...

Страница 149: ......

Страница 150: ...o not include a decimal point Resistance measurement values consist of four digits with FAST sampling Numerical values do not include a decimal point MEASure RESistance Measured resistance Judgment result FAIL PASS OFF NG ΩV HI IN LO OFF NG Ω Resistance measurement values consist of five digits with FAST sampling Numerical values do not include a decimal point Resistance measurement values consist...

Страница 151: ......

Страница 152: ......

Страница 153: ......

Страница 154: ...utput INIT CONT OFF vbCrLf Continuous measurement OFF For i 1 To 10 Wait for PC key input Create a key input check routine to set InputKey True when a key is pressed Do While 1 If InputKey True Then Exit Do DoEvents Loop After confirming key input measure once and read the measured value MSComm1 Output READ vbCrLf Send READ to measure and import the measurement recvstr From here on continue receiv...

Страница 155: ......

Страница 156: ...nal triggering MSComm1 Output INIT CONT ON vbCrLf Continuous measurement ON Clear confirmation of External I O TRIG input MSComm1 Output IO IN vbCrLf recvstr While Right recvstr 1 Chr 10 recvstr recvstr MSComm1 Input DoEvents Wend For i 1 To 10 Wait for External I O TRIG input Do While 1 MSComm1 Output IO IN vbCrLf recvstr While Right recvstr 1 Chr 10 recvstr recvstr MSComm1 Input DoEvents Wend If...

Страница 157: ......

Страница 158: ...ate unused Dim i As Integer pad 0 Board Address 0 gpibad 1 3561 3561 01 Address 1 timeout T10s Timeout about 10s Call ibfind gpib0 0 Initialize GP IB Call ibdev pad gpibad 0 timeout 1 0 ud Call SendIFC pad Open App Path data csv For Output As 1 Open a text file for saving Call Send pad gpibad TRIG SOUR IMM NLend Select internal triggering Call Send pad gpibad INIT CONT ON NLend Continuous measurem...

Страница 159: ......

Страница 160: ...ad 0 Board Address 0 gpibad 1 3561 3561 01 Address 1 timeout T100s Timeout 100s because of external trigger wait state Call ibfind gpib0 0 Initialize GP IB Call ibdev pad gpibad 0 timeout 1 0 ud Call SendIFC pad Open App Path data csv For Output As 1 Open a text file for saving Call Send pad gpibad TRIG SOUR EXT NLend Select external triggering Call Send pad gpibad INIT CONT OFF NLend Continuous m...

Страница 161: ......

Страница 162: ...v pad gpibad 0 timeout 1 0 ud Call SendIFC pad Call Send pad gpibad FUNC RV NLend Select ΩV mode Call Send pad gpibad RES RANG 300E 3 NLend Select 300 mΩ range Call Send pad gpibad SAMP RATE SLOW NLend Select SLOW sampling Call Send pad gpibad TRIG SOUR IMM NLend Select internal triggering Call Send pad gpibad INIT CONT OFF NLend Continuous measurement OFF Call Send pad gpibad CALC LIM BEEP HL NLe...

Страница 163: ......

Страница 164: ... becomes stable within the measurement accuracy of the internal measurement circuitry Resistance measurement Approx 3 ms Voltage measurement Approx 3 ms Response times are nominal values Actual values depend on the impedance characteristics of the object being measured Total measurement time Overall time required for measurement Response time sampling time Zero Adjustment Zero adjustment function ...

Страница 165: ......

Страница 166: ...I O Input CMOS level Output Open collector 35 V 50 mA max Input signals Measurement start trigger print zero adjustment calibration manual comparator and panel load 7 bit Output signals End of measurement End measurement Comparator result resistance Hi IN Lo voltage Hi IN L AND measurement fault and General Purpose output 10 bit EXT I O control input can be disabled by a remote command RS 232C Com...

Страница 167: ......

Страница 168: ......

Страница 169: ......

Страница 170: ......

Страница 171: ......

Страница 172: ......

Страница 173: ......

Страница 174: ......

Страница 175: ......

Страница 176: ......

Страница 177: ......

Страница 178: ......

Страница 179: ......

Страница 180: ......

Страница 181: ......

Страница 182: ...Appendix 9 Dimensional Diagram 176 ...

Страница 183: ......

Страница 184: ...ge 26 Measurement setting state 61 Measurement settings 62 Measurement time 76 Measurement value output 64 Memory function 58 128 Menu display 13 N Noise 4 6 132 163 166 NR1 95 NR2 95 NR3 95 O OF 32 157 Operating key 9 11 65 Option 1 172 OUT 72 Output queue 97 Overflow 32 157 P Panel load function 62 Panel save function 61 Period of guaranteed accuracy 162 Photocoupler connection 77 78 Power cord ...

Страница 185: ...ical calculation function 54 Sub display 9 10 Supply frequency 162 Supply voltage 6 19 162 Switch connection 77 Switching 50 Switching between measurement value and comparator setting displays 50 T Temperature coefficient 161 Terminator 94 Test lead 17 21 28 172 Threshold method 45 Timing chart 75 TRIG 51 56 58 64 71 Trigger 11 51 135 Trigger delay 11 52 Trigger source 135 U Upper and lower thresh...

Страница 186: ...iv Index ...

Страница 187: ......

Страница 188: ......

Страница 189: ...81 268 28 0568 E mail os com hioki co jp URL http www hioki co jp Printed in Japan 3561A981 01 All reasonable care has been taken in the production of this manual but if you find any points which are unclear or in error please contact your supplier or the International Sales and Mar keting Department at HIOKI headquarters In the interests of product development the contents of this manual are subj...

Страница 190: ...92 Japan TEL 81 268 28 0562 FAX 81 268 28 0568 E mail os com hioki co jp URL http www hioki co jp HIOKI USA CORPORATION 6 Corporate Drive Cranbury NJ 08512 USA TEL 1 609 409 9109 FAX 1 609 409 9108 3561A981 01 06 09H Printed on recycled paper ...

Отзывы: