3.6 Battery Measurement
50
Remove the test lead from the battery and turn off
the power to the instrument by pressing
but-
ton.
Stabilizing
measurement values
Storing measurement
values
Measurement values can be
stored by pressing
key
while the values are being
held.(
Loading stored data to
a personal computer
Configuring threshold
values and judging bat-
tery wear
Based on these wear judg-
ment values, threshold values
can be configured for judging
battery wear.(
• A measurement value display of "OF" indicates
that the resistance or voltage values displayed
exceed the relevant measurement range.
• If the voltage input is "OF," a buzzer will sound.
• A resistance value display of "- - - -" indicates
that the test lead is open or measurement cannot
be conducted due to an electrical current irregu-
larity caused by a disconnected test lead or other
reasons. "- - - -" will also be displayed when the
test lead is not correctly connected to the subject
of measurement or the subject's resistance is
substantially greater than the measurement
range.
• When measuring resistance of a relay or a con-
nector, note the maximum open circuit voltage of
this instrument (approximately 5 V). There is a
possibility that such measurement could damage
the connectors or oxidized coating of the subject
of measurement, leading to incorrect measure-
ment.
Ending the Measurement
Содержание 3554
Страница 1: ...BATTERY HiTESTER INSTRUCTION MANUAL 3554 ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 20: ...12 ...
Страница 32: ...1 6 External Dimensions 24 1 6 External Dimensions 192 1 mm 7 56 0 04 121 1 mm 4 76 0 04 55 1 mm 2 17 0 04 ...
Страница 80: ...5 5 Deleting Saved Data 72 ...
Страница 116: ...7 4 Using the Software 108 3 Click New File Edit Comparator Table will appear Click ...
Страница 132: ...7 4 Using the Software 124 ...
Страница 165: ......
Страница 166: ......