To ensure safe measurement, be sure to read
this section prior to measuring.
Measurement
Chapter 3
Observe the following to avoid electric shock
and short circuits.
• Do not measure voltages of 60 V DC or
higher.
• Do not measure grounded voltages of more
than 70 V DC.
• Do not measure alternating voltage.
Maximum rated voltage
to earth: 70 V DC
Maximum input voltage:
60 V DC
• Note that the instrument may be damaged if the
applied voltage or current exceeds the mea-
surement range.
• After measuring a high-voltage battery, before
continuing to measure a low-voltage battery first
short the test leads together. This will discharge
the DC elimination capacitor which is connected
across the leads. Otherwise an excess voltage
may be applied to the low-voltage battery, witch
is may cause damage.