![Hioki 3455 Скачать руководство пользователя страница 87](http://html.mh-extra.com/html/hioki/3455/3455_instruction-manual_2137473087.webp)
4.2 Displaying PI and DAR
1
2
3
4
5
6
7
8
9
10
11
付録
索引
83
• The tester automatically calculates and
displays PI (polarization index) and DAR
(dielectric absorption ratio), which are
used as the criteria to determine the qual-
ity of insulation.
Both measurements show a degree of
change in insulation resistance with time
after a test voltage is applied.
❖
Appendix 3 "Example of PI Criteria (Polarization
Index)" (page 170)
• PI and DAR are calculated using the for-
mulae below from resistance values mea-
sured twice after a voltage is applied. For
PI, the measurement interval may be
user-set.
❖
See 6.1 "Changing and Checking Interval Set-
ting for PI Calculation" (page 123).
Procedure
4.2 Displaying PI and DAR
What is it used for?
Used to check whether insulation re-
sistance increases with time after a
voltage is applied.
(When the PI value or the DAR val-
ue is close to 1, the tester deter-
mines that the insulation of the
object under test is deteriorated.)
PI 10/1min
=
DAR 1min/15s =
DAR 1min/30s =
To determine DAR, press the
key to
turn off [
AVE
] on the screen before starting
measurement.
Resistance 10 min. after voltage application
Resistance 1 min. after voltage application
Resistance 1 min. after voltage application
Resistance 15 sec. after voltage application
Resistance 1 min. after voltage application
Resistance 30 sec. after voltage application