![Hioki 3455 Скачать руководство пользователя страница 59](http://html.mh-extra.com/html/hioki/3455/3455_instruction-manual_2137473059.webp)
3.1 Pre-Operation Inspection
1
2
3
4
5
6
7
8
9
10
11
付録
索引
55
7.
Check to see if the insulation resistance
reading of the tester is 20 M
Ω
.
8.
Stop insulation resistance measurement.
❖
See 3.2.2 "Ending Measurement" (page 64).
9.
Short-circuit the tips of the clips of the
red and black test leads of the tester.
10.
Press the
key to see if the test volt-
age setting is [
5.00 kV
].
11.
Hold down the
key for more than
one second to start insulation resistance
measurement.
12.
Check to see if the insulation resistance
reading of the tester is 0.00 M
Ω
.
If a problem exists, discontinue use of
the tester.
Insulation resistance
Voltage
www.
.com