8.3 Error Display
1
2
3
4
5
6
7
8
9
10
11
付録
索引
163
Err02
When power is turned on for
the first time after the backup
battery is replaced, Err02
appears.
Reinstall the
battery.
If Err02 appears even after
the battery has been rein-
stalled, the life of the backup
battery has expired, the bat-
tery is faulty, or some other
cause exists.
This
requires
repair.
(Err03 to
Err05 may
be tempo-
rarily dis-
played
during dis-
charging
after mea-
surement,
but this does
not indicate
a malfunc-
tion.)
Err03
Voltage measurement error
has occurred.
Err04
Current measurement error
has occurred.
Err05
Temperature measurement
error has occurred.
Err06
The discharge circuit is faulty.
Err10
Details:
A high residual voltage is generated in an
object under insulation resistance testing.
High residual voltages generated in an
object placed under high test voltage condi-
tions results in charge being stored, and
then released under a subsequent test
involving lower test voltages.
Action:
• Stop measurement and discharge the
charge stored in the object under test
before resuming measurement.
• Even if the voltage has been lowered by
the tester’s automatic discharge function,
a high voltage may exist. Discharge using
a commercial discharging rod for high volt-
age.