Calibration Procedure
44
Exicor MicroImager User Manual
12. Tighten the screw as shown in Figure 27 to center the optic under the
objective, place blue mounting putty as shown in Figure 28 to keep the optic
from moving.
It is important that the optic be as flat as possible with
respect to the optical train.