For detailed Warning and Caution statements, and operating 
instructions, refer to the 

VersaCare® Bed User Manual

 (USR119  

for A through J models; 161956 for K model and newer).

NOTE:

 If the troubleshooting information shown 

below does not fix the problem, contact your 
facility-authorized maintenance person.

The Bed Does Not Work

Make sure the bed is plugged into AC power or 
the battery is active. For battery information, 
refer to “Battery Back-up” on page 4.

The Service Is Required (Wrench) 

Indicator Is On or Flashing

When the system determines the bed or treat-
ment/therapy surface operation is not correct, 
the wrench indicator comes on solid (bed error) 
or flashes (treatment/therapy surface error). 
Contact your facility-authorized maintenance 
person.

The Bed Controls Do Not Work

• Make sure the bed is plugged into AC power 

or the battery is active.

• Make sure the lockouts are deactivated. See 

page 6 to deactivate them.

NOTE: 

If all lockout indicators and the service 

required indicator are on, contact your facility-
authorized maintenance person.

The Bed Does Not Lower

Make sure of these:
• The bed is plugged into AC power or the bat-

tery is active.

• The Bed Down control is not locked out.
• The Bed Not Down indicator is not flashing.
• There is nothing between the upper frame 

and the base frame.

• There is nothing between the obstacle detect 

sensors on the base frame.

• The obstacle detect covers, located in the 

caster covers, are installed correctly.

The Foot Controls Do Not Work

Make sure of these:
• The bed is plugged into AC power 

or

 the bat-

tery is active.

• The lockouts are deactivated. 
• You have enabled the foot controls. Refer to 

“Foot Controls” on page 4.

The Display on the Control Pod Is Off

Make sure the bed is plugged into AC power. 
When the bed operates on battery power, the 
display will be off. 

NOTE: 

If the bed has the 30° Head Angle Alarm 

option and the alarm is set, if the head section 
goes below 30°, the audible alarm will come on 
and the Head Angle Alarm 

indicator

 will flash.

The Display on the Control Pod Flashes 
when a Weight Is Taken

The maximum weight the scale will show has 
been exceeded.

The Head Section Angle Appears to be 
much different than the Head Angle Dis-
play Shows (G model and newer beds)

Contact your facility-authorized maintenance 
person.

The Bed Exit Alarm Does Not Arm and 
All Three Mode Indicators Are Flashing

Remove the patient, and zero the Bed Exit sys-
tem. See “Bed Exit Alarm System” on page 8.

Troubleshooting

Bed

Содержание VersaCare

Страница 1: ...VersaCare Bed Quick Reference Guide...

Страница 2: ...ontrols 4 D Point of Care Caregiver Control Panel 5 6 E Control Pod 7 8 9 F IntelliDrive Transport System optional 10 G Active Integrated Response Treatment Surface or P500 Surface optional 12 H SafeV...

Страница 3: ...either AC or battery for the Emergency Trendelenburg function to operate 1 Make sure the bed is plugged into AC power or the battery is activated 2 Pull and hold the CPR handle 3 Hold the CPR handle u...

Страница 4: ...flash Battery Back up When the bed does not receive AC power bat tery back up can be activated When the bed operates on battery back up The functions on the caregiver control panel are available The...

Страница 5: ...r each time you press another control Caregiver Control Panel Controls Chair 1 Set the brake 2 A through F models press the Enable control G model and newer go to step 3 3 Press and hold the Chair con...

Страница 6: ...the bed to go into the 7 Tren delenburg position 2 Move the patient to the head of the bed as necessary 3 To deactivate the Boost function press any of the bed function controls except these Boost Sys...

Страница 7: ...atient is lying still and is centered on the mattress 1 A through F models press the Enable control G model and newer go to step 2 2 Press the Weigh con trol The bed shows the current patient weight o...

Страница 8: ...n The system has malfunctioned NOTE If a Bed Exit mode is active and the bed gets unplugged or the battery activated the Bed Exit Alarm System will be inactive When the bed is plugged in again the Bed...

Страница 9: ...head angle alarm is active and the head section goes below the specified angle Bed Exit is in Silence mode bed with the Bed Exit Alarm Silence fea ture Alerts Flash Yellow and Green there is a techni...

Страница 10: ...ed 7 Press one or both of the Enable switches on the transport handles 8 While you press the Enable switches push the transport handles forward or pull them backward for the applicable direction 9 To...

Страница 11: ...er or the bat tery is active The Bed Down control is not locked out The Bed Not Down indicator is not flashing There is nothing between the upper frame and the base frame There is nothing between the...

Страница 12: ...become very firm After 30 minutes if no other mode is selected the surface automatically returns to the Pressure Relief Normal mode NOTE If the surface is in Max Inflate mode and the bed gets unplugge...

Страница 13: ...e siderails are not in the correct position or the head section is not below 25 an indicator will flash and an alarm will sound Deactivate 1 Press the Enable con trol 2 Press the opposite Turn control...

Страница 14: ...solid To activate Sleep mode refer to Sleep Mode on page 13 The scale system is used correctly If it is not used correctly the surface may go into the non operational safe mode see All Surface Mode I...

Страница 15: ...the sleep deck Use applicable precautions with gravity sensitive devices such as ventricular drains or lumbar drains before during and after you operate the bed functions Make sure flow rates on all g...

Страница 16: ...ween 57 and 84 145 cm to 213 cm 2011 by Hill Rom Services Inc ALL RIGHTS RESERVED No part of this text shall be reproduced without written permission from Hill Rom Services Inc Active Integrated Respo...

Отзывы: