For detailed Warning and Caution statements, and operating
instructions, refer to the
VersaCare® Bed User Manual
(USR119
for A through J models; 161956 for K model and newer).
NOTE:
If the troubleshooting information shown
below does not fix the problem, contact your
facility-authorized maintenance person.
The Bed Does Not Work
Make sure the bed is plugged into AC power or
the battery is active. For battery information,
refer to “Battery Back-up” on page 4.
The Service Is Required (Wrench)
Indicator Is On or Flashing
When the system determines the bed or treat-
ment/therapy surface operation is not correct,
the wrench indicator comes on solid (bed error)
or flashes (treatment/therapy surface error).
Contact your facility-authorized maintenance
person.
The Bed Controls Do Not Work
• Make sure the bed is plugged into AC power
or the battery is active.
• Make sure the lockouts are deactivated. See
page 6 to deactivate them.
NOTE:
If all lockout indicators and the service
required indicator are on, contact your facility-
authorized maintenance person.
The Bed Does Not Lower
Make sure of these:
• The bed is plugged into AC power or the bat-
tery is active.
• The Bed Down control is not locked out.
• The Bed Not Down indicator is not flashing.
• There is nothing between the upper frame
and the base frame.
• There is nothing between the obstacle detect
sensors on the base frame.
• The obstacle detect covers, located in the
caster covers, are installed correctly.
The Foot Controls Do Not Work
Make sure of these:
• The bed is plugged into AC power
or
the bat-
tery is active.
• The lockouts are deactivated.
• You have enabled the foot controls. Refer to
The Display on the Control Pod Is Off
Make sure the bed is plugged into AC power.
When the bed operates on battery power, the
display will be off.
NOTE:
If the bed has the 30° Head Angle Alarm
option and the alarm is set, if the head section
goes below 30°, the audible alarm will come on
and the Head Angle Alarm
indicator
will flash.
The Display on the Control Pod Flashes
when a Weight Is Taken
The maximum weight the scale will show has
been exceeded.
The Head Section Angle Appears to be
much different than the Head Angle Dis-
play Shows (G model and newer beds)
Contact your facility-authorized maintenance
person.
The Bed Exit Alarm Does Not Arm and
All Three Mode Indicators Are Flashing
Remove the patient, and zero the Bed Exit sys-
tem. See “Bed Exit Alarm System” on page 8.