For detailed Warning and Caution statements, and operating 
instructions, refer to the 

VersaCare® Bed User Manual

 (USR119  

for A through J models; 161956 for K model and newer).

Bed

Obstacle Detection 

and Battery and Foot 

Controls

Obstacle Detection

The bed is equipped with an obstacle detection 
system that runs along the three open sides of 
the base frame. This system detects objects 
that are between the upper frame and the base 
frame. 
If an object is detected, the Bed Not Down indi-
cator on both siderails will flash and you will not 
be able to lower the bed. 
If an object is detected while the bed sleep deck 
is lowering, the bed will stop lowering, and then 
raise automatically for 2 seconds. The Bed Not 
Down indicator on both siderails will flash.

Battery Back-up

When the bed does not receive AC power, bat-
tery back-up can be activated. When the bed 
operates on battery back-up—
•  The functions on the caregiver control panel 

are

 available.

• The functions on the control pod are 

not

 

available, and the surface indicators are off.

• The battery stays active for 10 minutes (A 

through J models) or 1 minute (K model and 
newer) after the last control is used.

The battery indicator on the caregiver control 
panel shows its status (see “Battery” on page 3).

Activate 

A through J models

—press the Battery con-

trol that is at the right-side head end, just 
below the power cord. 

K model and newer

—press 

any

 caregiver 

siderail control except the Lockout control. 

Foot Controls

1. To 

enable

 the controls, step down on one of 

the 

Head Down/Up 

or

 Bed Down/Up

 ped-

als for 1 to 3 seconds.

2. Release the pedal.
3. Step down on the pedal of the applicable 

function until the bed is in the correct posi-
tion.

Bed Down/Up

Head Down/Up

Содержание VersaCare

Страница 1: ...VersaCare Bed Quick Reference Guide...

Страница 2: ...ontrols 4 D Point of Care Caregiver Control Panel 5 6 E Control Pod 7 8 9 F IntelliDrive Transport System optional 10 G Active Integrated Response Treatment Surface or P500 Surface optional 12 H SafeV...

Страница 3: ...either AC or battery for the Emergency Trendelenburg function to operate 1 Make sure the bed is plugged into AC power or the battery is activated 2 Pull and hold the CPR handle 3 Hold the CPR handle u...

Страница 4: ...flash Battery Back up When the bed does not receive AC power bat tery back up can be activated When the bed operates on battery back up The functions on the caregiver control panel are available The...

Страница 5: ...r each time you press another control Caregiver Control Panel Controls Chair 1 Set the brake 2 A through F models press the Enable control G model and newer go to step 3 3 Press and hold the Chair con...

Страница 6: ...the bed to go into the 7 Tren delenburg position 2 Move the patient to the head of the bed as necessary 3 To deactivate the Boost function press any of the bed function controls except these Boost Sys...

Страница 7: ...atient is lying still and is centered on the mattress 1 A through F models press the Enable control G model and newer go to step 2 2 Press the Weigh con trol The bed shows the current patient weight o...

Страница 8: ...n The system has malfunctioned NOTE If a Bed Exit mode is active and the bed gets unplugged or the battery activated the Bed Exit Alarm System will be inactive When the bed is plugged in again the Bed...

Страница 9: ...head angle alarm is active and the head section goes below the specified angle Bed Exit is in Silence mode bed with the Bed Exit Alarm Silence fea ture Alerts Flash Yellow and Green there is a techni...

Страница 10: ...ed 7 Press one or both of the Enable switches on the transport handles 8 While you press the Enable switches push the transport handles forward or pull them backward for the applicable direction 9 To...

Страница 11: ...er or the bat tery is active The Bed Down control is not locked out The Bed Not Down indicator is not flashing There is nothing between the upper frame and the base frame There is nothing between the...

Страница 12: ...become very firm After 30 minutes if no other mode is selected the surface automatically returns to the Pressure Relief Normal mode NOTE If the surface is in Max Inflate mode and the bed gets unplugge...

Страница 13: ...e siderails are not in the correct position or the head section is not below 25 an indicator will flash and an alarm will sound Deactivate 1 Press the Enable con trol 2 Press the opposite Turn control...

Страница 14: ...solid To activate Sleep mode refer to Sleep Mode on page 13 The scale system is used correctly If it is not used correctly the surface may go into the non operational safe mode see All Surface Mode I...

Страница 15: ...the sleep deck Use applicable precautions with gravity sensitive devices such as ventricular drains or lumbar drains before during and after you operate the bed functions Make sure flow rates on all g...

Страница 16: ...ween 57 and 84 145 cm to 213 cm 2011 by Hill Rom Services Inc ALL RIGHTS RESERVED No part of this text shall be reproduced without written permission from Hill Rom Services Inc Active Integrated Respo...

Отзывы: