For detailed Warning and Caution statements, and operating 
instructions, refer to the 

VersaCare® Bed User Manual

 (USR119  

for A through J models; 161956 for K model and newer).

Emergency CPR and 

Trendelenburg and 

Safety and Information 

Indicators

Bed

 

Emergency CPR

The emergency CPR controls are handles 
located under the sleep deck, between the head 
and intermediate siderails on both sides of the 
bed.

NOTE: 

The headboard can be used as a CPR 

board.
1. Pull and hold the 

CPR 

handle. If the bed has 

an treatment/therapy surface, the surface 
will go into Max-Inflate.

2. Hold the 

CPR 

handle until the head and 

knee section come to a stop in the flat posi-
tion. 

3. Release the 

CPR 

handle.

Emergency Trendelenburg

The emergency Trendelenburg controls are 
handles located under the sleep deck, between 
the head and intermediate siderails on both 
sides of the bed. The handles used to operate 
CPR are also used to operate emergency Tren-
delenburg.

NOTE: 

There must be power to the bed, either 

AC or battery, for the Emergency Trendelenburg 
function to operate.
1. Make sure the bed is plugged into AC power 

or the battery is activated.

2. Pull and hold the 

CPR 

handle.

3. Hold the 

CPR 

handle until the head and 

knee section come to a stop in the flat posi-
tion.

4. Continue to hold the 

CPR 

handle until you 

reach the applicable bed angle.

5. Release the 

CPR 

handle.

Safety and Information 
Indicators 

NOTE: 

For the indicators to oper-

ate, there must be AC or battery 
power to the bed. 

NOTE:

 

A through F models

 have the wrench 

(Service is Required) indicator 

only

.

Plug (Disconnected from AC Power)

If the bed is disconnected from AC power and 
the battery is active, the plug indicator flashes.

Wrench (Service Is Required) 

When the wrench indicator comes on solid or 
flashes, contact your facility-authorized mainte-
nance person.

Battery

• ON = battery is active.
• FLASHING = battery needs to be charged.
• OFF = battery is inactive or is discharged 

below the level necessary to operate the 
motors.

Brake Not Set 

If the bed is connected to AC power, and the 
brake is not set, the Brake Not Set indicator 
(available on some beds) flashes and a continu-
ous alarm comes on. Set the brake to stop the 
alarm.

Содержание VersaCare

Страница 1: ...VersaCare Bed Quick Reference Guide...

Страница 2: ...ontrols 4 D Point of Care Caregiver Control Panel 5 6 E Control Pod 7 8 9 F IntelliDrive Transport System optional 10 G Active Integrated Response Treatment Surface or P500 Surface optional 12 H SafeV...

Страница 3: ...either AC or battery for the Emergency Trendelenburg function to operate 1 Make sure the bed is plugged into AC power or the battery is activated 2 Pull and hold the CPR handle 3 Hold the CPR handle u...

Страница 4: ...flash Battery Back up When the bed does not receive AC power bat tery back up can be activated When the bed operates on battery back up The functions on the caregiver control panel are available The...

Страница 5: ...r each time you press another control Caregiver Control Panel Controls Chair 1 Set the brake 2 A through F models press the Enable control G model and newer go to step 3 3 Press and hold the Chair con...

Страница 6: ...the bed to go into the 7 Tren delenburg position 2 Move the patient to the head of the bed as necessary 3 To deactivate the Boost function press any of the bed function controls except these Boost Sys...

Страница 7: ...atient is lying still and is centered on the mattress 1 A through F models press the Enable control G model and newer go to step 2 2 Press the Weigh con trol The bed shows the current patient weight o...

Страница 8: ...n The system has malfunctioned NOTE If a Bed Exit mode is active and the bed gets unplugged or the battery activated the Bed Exit Alarm System will be inactive When the bed is plugged in again the Bed...

Страница 9: ...head angle alarm is active and the head section goes below the specified angle Bed Exit is in Silence mode bed with the Bed Exit Alarm Silence fea ture Alerts Flash Yellow and Green there is a techni...

Страница 10: ...ed 7 Press one or both of the Enable switches on the transport handles 8 While you press the Enable switches push the transport handles forward or pull them backward for the applicable direction 9 To...

Страница 11: ...er or the bat tery is active The Bed Down control is not locked out The Bed Not Down indicator is not flashing There is nothing between the upper frame and the base frame There is nothing between the...

Страница 12: ...become very firm After 30 minutes if no other mode is selected the surface automatically returns to the Pressure Relief Normal mode NOTE If the surface is in Max Inflate mode and the bed gets unplugge...

Страница 13: ...e siderails are not in the correct position or the head section is not below 25 an indicator will flash and an alarm will sound Deactivate 1 Press the Enable con trol 2 Press the opposite Turn control...

Страница 14: ...solid To activate Sleep mode refer to Sleep Mode on page 13 The scale system is used correctly If it is not used correctly the surface may go into the non operational safe mode see All Surface Mode I...

Страница 15: ...the sleep deck Use applicable precautions with gravity sensitive devices such as ventricular drains or lumbar drains before during and after you operate the bed functions Make sure flow rates on all g...

Страница 16: ...ween 57 and 84 145 cm to 213 cm 2011 by Hill Rom Services Inc ALL RIGHTS RESERVED No part of this text shall be reproduced without written permission from Hill Rom Services Inc Active Integrated Respo...

Отзывы: