Page 9 of 16
For detailed Warning and Caution statements, and operating
instructions, refer to the
VersaCare® Bed User Manual
(USR119
for A through J models; 161956 for K model and newer).
SafeView™ Alerts
and 30° Head Angle
Alarm
Bed
SafeView™ Alerts
When the bed is connected to AC power and the Bed
Exit System is active, the Alerts show the safety con-
dition of the bed as defined:
• Bed is in the low position.
• Siderails are up (as configured).
• Brake is set.
NOTE
: If the head angle alarm is installed and active, it is included in the
safe condition requirements.
Alerts Are On Green
—Bed Exit is active, and the conditions shown
above are met.
Alerts Flash Yellow
—
• Bed Exit is active, and one or more of the conditions shown above are
not met.
• Bed Exit is not active, the optional head angle alarm is active, and the
head section goes below the specified angle.
• Bed Exit is in Silence mode (bed with the Bed Exit Alarm Silence fea-
ture).
Alerts Flash Yellow and Green
—there is a technical problem with the
Alerts. Call your facility-authorized maintenance person.
Alerts are Off
—Bed Exit is not active, the Alerts have been deactivated,
or the bed is disconnected from AC power.
30° Head Angle Alarm (G Model and Newer)
Activate
1. Raise the head section above 30°.
2. Press the
Enable
control.
3. Press the
Alarm
control until its indicator is on solid. The
indicator stays on until the alarm is deactivated.
NOTE:
When the bed operates on battery power and the
alarm is set, if the head section goes below 30°, an audible
alarm will come on and the alarm indicator will flash. The dis-
play on the control pod will be off.
Turn Off the Alarm
Raise the head section above 30°.
Deactivate
1. Press the
Enable
control.
2. Press the
Alarm
control until its indicator is off.