For detailed Warning and Caution statements, and operating
instructions, refer to the
VersaCare® Bed User Manual
(USR119
for A through J models; 161956 for K model and newer).
Bed Exit Alarm System
Activate
1. Make sure the patient is
not
on the bed.
2. Zero the scale. See “Scale” on page 7.
3. Put the patient in the center of the bed and
aligned with the hip locator.
4. Press the
Enable
control.
5. Press the applicable
Bed Exit
mode control.
When the system beeps one time and the
indicator stays on solid, the system is armed.
If the system does not arm, the system will beep
rapidly for a few seconds and the selected
mode indicator will flash. This means:
• The patient weight is outside these ranges:
A model
—70 lb to 500 lb (31.8 kg to 227 kg)
B model and newer
—50 lb to 500 lb (23 kg
to 227 kg)
• The patient is not in the correct position.
• The system has malfunctioned.
NOTE:
If a Bed Exit mode is active and the bed
gets unplugged or the battery activated, the Bed
Exit Alarm System will be
inactive
. When the
bed is plugged in again, the Bed Exit mode will
activate at the previous setting.
NOTE:
When there is a Bed Exit alarm condi-
tion, the mode indicator flashes and an alarm
comes on.
Deactivate
1. Press the
Enable
control.
2. Press any Bed Exit mode control until its
indicator is off.
Turn Off the Alarm Sound
•
Bed without the Alarm Silence control
—
you must deactivate the Bed Exit Alarm Sys-
tem.
•
Bed with the Alarm Silence control
—either
deactivate the Bed Exit Alarm System, or use
the Alarm Silence control to temporarily turn
off the alarm sound.
Alarm Silence
While Bed Exit is alarming,
press the
Alarm Silence
control until its indicator is
on solid.
The system stays in Silence mode until the
patient is on the bed and has met all the condi-
tions to not cause an alarm for the current Bed
Exit setting.
Patient Position
Out-of-Bed
Bed Exiting
Bed
Bed Exit Alarm System