Page 8 of 16 

For detailed Warning and Caution statements, and operating 
instructions, refer to the 

VersaCare® Bed User Manual

 (USR119  

for A through J models; 161956 for K model and newer).

Bed Exit Alarm System

Activate

1. Make sure the patient is 

not

 on the bed.

2. Zero the scale. See “Scale” on page 7.
3. Put the patient in the center of the bed and 

aligned with the hip locator.

4. Press the 

Enable

 control.

5. Press the applicable 

Bed Exit

 mode control. 

When the system beeps one time and the 
indicator stays on solid, the system is armed.

If the system does not arm, the system will beep 
rapidly for a few seconds and the selected 
mode indicator will flash. This means: 

• The patient weight is outside these ranges: 

A model

—70 lb to 500 lb (31.8 kg to 227 kg) 

B model and newer

—50 lb to 500 lb (23 kg 

to 227 kg)

• The patient is not in the correct position.
• The system has malfunctioned.

NOTE:

 If a Bed Exit mode is active and the bed 

gets unplugged or the battery activated, the Bed 
Exit Alarm System will be 

inactive

. When the 

bed is plugged in again, the Bed Exit mode will 
activate at the previous setting.

NOTE:

 When there is a Bed Exit alarm condi-

tion, the mode indicator flashes and an alarm 
comes on. 

Deactivate

1. Press the 

Enable

 control.

2. Press any Bed Exit mode control until its 

indicator is off.

Turn Off the Alarm Sound

Bed without the Alarm Silence control

you must deactivate the Bed Exit Alarm Sys-
tem.

Bed with the Alarm Silence control

—either 

deactivate the Bed Exit Alarm System, or use 
the Alarm Silence control to temporarily turn 
off the alarm sound.

Alarm Silence

While Bed Exit is alarming, 
press the 

Alarm Silence

 

control until its indicator is 
on solid. 
The system stays in Silence mode until the 
patient is on the bed and has met all the condi-
tions to not cause an alarm for the current Bed 
Exit setting. 

Patient Position

Out-of-Bed

Bed Exiting

Bed

Bed Exit Alarm System

Содержание VersaCare

Страница 1: ...VersaCare Bed Quick Reference Guide...

Страница 2: ...ontrols 4 D Point of Care Caregiver Control Panel 5 6 E Control Pod 7 8 9 F IntelliDrive Transport System optional 10 G Active Integrated Response Treatment Surface or P500 Surface optional 12 H SafeV...

Страница 3: ...either AC or battery for the Emergency Trendelenburg function to operate 1 Make sure the bed is plugged into AC power or the battery is activated 2 Pull and hold the CPR handle 3 Hold the CPR handle u...

Страница 4: ...flash Battery Back up When the bed does not receive AC power bat tery back up can be activated When the bed operates on battery back up The functions on the caregiver control panel are available The...

Страница 5: ...r each time you press another control Caregiver Control Panel Controls Chair 1 Set the brake 2 A through F models press the Enable control G model and newer go to step 3 3 Press and hold the Chair con...

Страница 6: ...the bed to go into the 7 Tren delenburg position 2 Move the patient to the head of the bed as necessary 3 To deactivate the Boost function press any of the bed function controls except these Boost Sys...

Страница 7: ...atient is lying still and is centered on the mattress 1 A through F models press the Enable control G model and newer go to step 2 2 Press the Weigh con trol The bed shows the current patient weight o...

Страница 8: ...n The system has malfunctioned NOTE If a Bed Exit mode is active and the bed gets unplugged or the battery activated the Bed Exit Alarm System will be inactive When the bed is plugged in again the Bed...

Страница 9: ...head angle alarm is active and the head section goes below the specified angle Bed Exit is in Silence mode bed with the Bed Exit Alarm Silence fea ture Alerts Flash Yellow and Green there is a techni...

Страница 10: ...ed 7 Press one or both of the Enable switches on the transport handles 8 While you press the Enable switches push the transport handles forward or pull them backward for the applicable direction 9 To...

Страница 11: ...er or the bat tery is active The Bed Down control is not locked out The Bed Not Down indicator is not flashing There is nothing between the upper frame and the base frame There is nothing between the...

Страница 12: ...become very firm After 30 minutes if no other mode is selected the surface automatically returns to the Pressure Relief Normal mode NOTE If the surface is in Max Inflate mode and the bed gets unplugge...

Страница 13: ...e siderails are not in the correct position or the head section is not below 25 an indicator will flash and an alarm will sound Deactivate 1 Press the Enable con trol 2 Press the opposite Turn control...

Страница 14: ...solid To activate Sleep mode refer to Sleep Mode on page 13 The scale system is used correctly If it is not used correctly the surface may go into the non operational safe mode see All Surface Mode I...

Страница 15: ...the sleep deck Use applicable precautions with gravity sensitive devices such as ventricular drains or lumbar drains before during and after you operate the bed functions Make sure flow rates on all g...

Страница 16: ...ween 57 and 84 145 cm to 213 cm 2011 by Hill Rom Services Inc ALL RIGHTS RESERVED No part of this text shall be reproduced without written permission from Hill Rom Services Inc Active Integrated Respo...

Отзывы: