Preventive Maintenance (Capital Products Only)
Chapter 6: General Procedures
Page 6 - 12
TotalCare® Bariatric Bed Service Manual (man299ra)
Preventive Maintenance Schedule and Checklist
Inspector Name: _______________________
Inspector Signature: ____________________
Inspector Employee Number: _____________
Date Inspected: _________________________
Bed Serial Number: ______________________ Model Number: _________________________
Labor Time: ____________________________ Facility Location: _______________________
Indicate each step as “Pass” or “Fail” in the right-hand column.
Pass/Fail
Manual Hydraulic Function Test
•
Unplug the bed from its power source, and verify proper operation of the head
functions.
Disinfect and Inspect The Power Cord
•
Inspect the length of the power cord for cuts and abrasions. Replace as needed.
•
Inspect the length of the pendant cord for cuts and abrasions. Replace as
needed.
•
Wipe down the entire length of the pendant cord and power cord with a facility-
approved disinfectant.
•
Inspect all head end warning labels. Replace labels as needed.
Ground Resistance Check
•
Plug the bed into a properly calibrated safety analyzer, and attach the safety
analyzer to the head end lift arm ground strap.
•
Check that electrical ground resistance readings do not exceed the maximum
specification of 200 milliohms.
Leakage Current Check
•
Using a calibrated safety analyzer, verify that the bed’s electrical leakage read-
ings meet the specification of 200 microamps.
G/NP __________ NG/NP _________ G/RP __________ NG/RP __________
•
Unplug the bed from the analyzer.
Disinfect and Inspect the Bed’s Head End
•
Plug the bed into an appropriate power source.
•
Raise the bed to its highest position.
•
Raise the head section 10°, place an angle locator on the head and confirm that
no greater than 3° from left to right exists.
•
Inspect for physical damage and missing parts, including the winch tab. Repair
or replace as necessary.
•
Inspect the IntelliDrive® Transport System handles for physical damage.
Repair or replace as necessary.
•
Verify the head end casters rotate 360° without touching the base frame.
Содержание TotalCare P1830A
Страница 2: ......
Страница 6: ...Revisions Page iv TotalCare Bariatric Bed Service Manual man299ra NOTES...
Страница 26: ...Chapter 2 Troubleshooting Procedures Page 2 2 TotalCare Bariatric Bed Service Manual man299ra NOTES...
Страница 30: ...Chapter 3 Theory of Operation Page 3 2 TotalCare Bariatric Bed Service Manual man299ra NOTES...
Страница 48: ...Chapter 5 Parts List Page 5 2 TotalCare Bariatric Bed Service Manual man299ra NOTES...
Страница 50: ...Warranty Chapter 5 Parts List Page 5 4 TotalCare Bariatric Bed Service Manual man299ra NOTES...
Страница 61: ...Head Siderail Module Chapter 5 Parts List TotalCare Bariatric Bed Service Manual man299ra Page 5 15 5 NOTES...
Страница 74: ...Chapter 6 General Procedures Page 6 2 TotalCare Bariatric Bed Service Manual man299ra NOTES...
Страница 92: ...Chapter 7 Accessories Page 7 2 TotalCare Bariatric Bed Service Manual man299ra NOTES...