Deflate the Mattress
1. Press Options. The Select Options
screen shows.
2. Press Deflate Bed. The Bed Deflate
Active screen shows, and the mattress
deflates.
3. To inflate the mattress, press Inflate
and Exit. The Bed Inflating screen
flashes for 10 seconds.
4. The unit goes into Hyper-Inflate mode
to inflate the mattress to 55 mm Hg
(29"H
2
O) for 13 minutes and then
returns to the previous therapy.
Solve a System Alarm Condition
If the unit detects an alarm condition in the
mattress, the System Alarm screen shows.
Solve a system alarm condition as follows:
1. Press Alarm Off to silence the alarm
for 10 minutes.
2. Refer to the System Alarm screen to
determine the alarm condition.
NOTE:
The System Alarm screen shows two rows of pressure in the three zones—head, seat, and
foot. The air cells are numbered #1 through #22, starting at the head end of the bed. To
determine which air cells and hose(s) cause the alarm condition, refer to the tables below.
Top Row Air Cell Connections to Hoses
Head
Seat
Foot
Hose color
Red
Blue
White
Air cells
1, 3, 5
7, 9, 11, 13
15, 17, 19
Bottom Row Air Cell Connections to Hoses
Head
Seat
Foot
Hose color
White
Black
Black
Air cells
2, 4, 6
8, 10, 12, 14
16, 18, 20
3. When the problem is solved, press Reset Alarm.
29