Max Firm
NOTE:
The mattress goes into Hyper-Inflate mode to inflate the mattress after Seat Deflate mode.
It does not goes into Max Firm mode.
For patient transfer and medical treatment, the Max Firm mode gives a firm surface. To
select Max Firm mode, do as follows:
1. Press Max Firm:
•
If the unit is in Low Airloss mode,
the Max Firm (65 mm Hg) screen
shows.
•
If unit is in Side Deflate mode, the
Side Deflate Active (Max Firm
Active) screen shows.
2. If the unit is in Max Firm mode for 30
minutes, the alarm sounds, and the
Warning: Max Firm Selected screen
shows. To turn the alarm off for 10
minutes, press Alarm Off.
3. To return to the previous therapy mode, press Max Firm again.
Fowler Boost
For patients in an inclined position, the
Fowler Boost therapy increases pressure in
the seat zone. To select the Fowler Boost,
press Fowler.
When the unit is in Fowler Boost mode, the
Fowler Boost indicator is on.
When the unit is turned on, it automatically defaults to Fowler Boost
mode. The standard Fowler Boost setting is 30% more than the seat
section’s set pressure. To change the Fowler Boost setting, do as follows:
1. Press Options. The Select Options
screen shows.
2. Press Adjust Fowler. The Fowler
Boost screen shows.
26