Set the Brakes or Steer the Bed
WARNING:
Unless transporting the patient, always set the brakes when the bed is occupied.
Reconfirm that the brakes are set before any patient transfer. Failure to do so may
result in personal injury or equipment damage.
The Brake and Steer controls are at the four corners of the bed
frame. There are three positions:
• Brake—keeps the bed from moving.
• Neutral—lets the caregiver move the bed sideways.
• Steer—lets the caregiver move the bed move in a straight
line.
Press the orange brake foot pedal down to lock the caster.
Press the green steer foot pedal down to unlock the caster.
Use the Battery Backup Control
The Battery control is at the right-hand corner, head end of the bed,
under the headboard just above the power cord.
To start battery backup power, press and hold the Battery control for
2 to 3 seconds. All articulation functions are available. The battery
stays active for 10 minutes after the last control is operated.
Charge the Battery
WARNING:
Keep the bed plugged into the power source until the battery indicator goes off
(charging time is approximately 10 hours). Failure to do so could result in the
inability to operate the bed when power is not available.
If the battery charge level is low, an indicator will flash. To recharge the battery, plug the
bed into an appropriate power source.
While the battery is charging, the indicator flashes. When the indicator stops flashing the
battery is fully charged. When plugged into an appropriate power source and the indicator
is not illuminated, the battery has lost all of its charge or is disconnected.