![Hill-Rom Centrella Smart+ Скачать руководство пользователя страница 27](http://html1.mh-extra.com/html/hill-rom/centrella-smart/centrella-smart_product-reference-manual_2134166027.webp)
27
©2017 Hill-Rom Services, Inc. ALL RIGHTS RESERVED.
205278 rev 2 13-SEP-2017 ENG – US
Scale Operations
Standard on all Centrella™ beds, the built-in scale allows you to take a
patient’s weight while in the bed.
Not included in Zero and Weigh Processes:
• Headboard and any device mounted to the head-end of the bed
Included in Zero and Weigh Processes:
• Equipment and linen attached to any part of the bed that moves, including
the siderails, footboard and foot-end sockets.
NOTES:
• Scale only operates when connected to power. Last zero and the history of
last weight are maintained when the bed is not connected to power.
• The scale is very sensitive. Make sure the bed is not touching anything that
could affect the patient weight, such as a wall or furniture. Failure to do so
could cause an inaccurate reading.
G R APH I C AL C AR EG I V E R I N T E R FACE (G CI)® TO U CHSCR E E N
Head Angle
32˚
0.0
<30˚
Limit
Stand
Assist