![Hill-Rom Centrella Smart+ Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/hill-rom/centrella-smart/centrella-smart_product-reference-manual_2134166022.webp)
22
©2017 Hill-Rom Services, Inc. ALL RIGHTS RESERVED.
205278 rev 2 13-SEP-2017 ENG – US
Stand Assist Feature
The Stand Assist control is accessed via the Patient pendant or from the GCI
Touchscreen. One-button simultaneously levels legs to a flat position and
raises the head-of-bed to position the patient to safely get out of the bed.
To activate, press and hold the Stand Assist control until the legs are flat and
the head is at the desired angle.
If equipped with a Centrella™ MAX powered air surface, wait for the seat
section to inflate to provide more support for the patient to get out of bed.
Finally, adjust the bed height until the patient’s feet are flat on the floor, and
help the patient out of the bed.
Bed Flat Control
The one-button Bed Flat and Level control, located on the GCI touchscreen,
returns the bed, head, knees and feet to the flat and level position from any
position.
To activate, press and hold the Bed Flat and Level control until the system
stops. You will hear a single beep when the bed is at its flat position.
NOTE: If any frame function is locked, the Bed Flat control will not operate, an
audible alert will sound and a message will appear on the touchscreen.
FlexAfoot™ Feature
The FlexAfoot™ bed length adjustment is a powered control standard on all
models. With the press of a button, the caregiver can adjust the foot section
up to 12” to accommodate various patient heights and make patients more
comfortable.
To lengthen, press and hold the Foot Longer control until foot section is at
the desired length.
To shorten, press and hold the Foot Shorter control until the desired length
is achieved.
NOTES:
• Length adjustment can be stopped anywhere between the fully retracted
(76” inner, 87” outer) and fully extended position (88” inner, 99” outer).
• The FlexAfoot™ feature is available on battery backup and can be used
during transport to adjust to small spaces such as an elevator.
G R APH I C AL C AR EG I V E R I N T E R FACE (G CI)® TO U CHSCR E E N
Head Angle
32˚
<30˚
Limit
Stand
Assist
Head Angle
32˚
<30˚
Limit
Stand
Assist
Head Angle
32˚
<30˚
Limit
Stand
Assist