Synergy® Air Elite Low Airloss Alternating Therapy System User Manual

 11

(139122 REV 1)

Mattress

To clean the mattress on the bed, use a damp soft cloth with mild 
detergent and warm water. Wipe down the air cells with a solution of 
sodium hypochlorite (1000 ppm) or any other non-phonetic germicidal 
solution.

If the top coverlet or bottom encasement is soiled, either discard it, or do 
as follows to machine wash and thermally disinfect it:

1. Put on applicable protective personal equipment (PPE).

Discard

a.

Remove and discard the top coverlet or bottom encasement in 
accordance with the facility’s procedures for contaminated 
waste.

b. Replace with a clean top coverlet or bottom encasement.

Wash

Industrial

 laundry—

a.

Break wash cycle in cold water for 10 minutes.

b. Main wash cycle I at 140°F (60°C) for 6 minutes.

c.

Main wash cycle II at 158°F (70°C) for 10 minutes.

d. Cold rinse for three standard cycles

e.

Extraction cycle for 5 minutes.

Normal

 laundry—

a.

Pre-wash in cold water for a standard cycle.

b. Main wash cycle at 158°F (70°C) for 10 minutes.

c.

Extraction cycle for 2 minutes.

d. Cold rinse for a standard cycle.

e.

Extraction cycle for 5 minutes.

2. Let the top coverlet or bottom encasement air dry, or machine dry at 

110°F (43°C) for one hour.

Steam Cleaning

Do not use any steam cleaning device on the unit. Excessive moisture 
can damage mechanisms in this unit.

Содержание careselections Synergy Air Elite

Страница 1: ...USER MANUAL Synergy Air Elite Low Airloss Alternating Therapy System From Hill Rom 139122 REV 1...

Страница 2: ......

Страница 3: ...l Rom is a registered trademark of Hill Rom Services Inc Synergy is a registered trademark of Mediq PRN Life Support Services Inc Underwriters Laboratories Inc is a registered trademark of Underwriter...

Страница 4: ...ii Synergy Air Elite Low Airloss Alternating Therapy System User Manual 139122 REV 1 Revision Pages Affected Date 1 December 2005...

Страница 5: ...ating Therapy Mode 8 Static Mode 9 Auto Firm Mode 9 Power Failure Alarm 9 Cleaning 10 General Cleaning 10 Control Unit 10 Mattress 11 Steam Cleaning 11 Cleaning Hard to Clean Spots 12 Disinfecting 12...

Страница 6: ...iv Synergy Air Elite Low Airloss Alternating Therapy System User Manual 139122 REV 1...

Страница 7: ...oldface text emphasizes a word or phrase NOTE sets apart special data or important instruction clarification WARNING or CAUTION A WARNING identifies situations or actions that may have an effect on pa...

Страница 8: ...ntrol Soft Firm Press the Soft or Firm control to adjust the Comfort Control setting shown on the eight increment scale Cycle Timer Press the Cycle Timer control to adjust the Alternating Therapy cycl...

Страница 9: ...95 minutes Control unit maximum flow rate 1200 lpm Control unit rated voltage 110 V to 120 V Control unit maximum current 5 A Control unit rated frequency 60 Hz Control unit protection type Class II...

Страница 10: ...the raised position Failure to do either of these could result in serious injury or death NOTE Siderails are intended to be a reminder to the patient of the unit s edges not a patient restraining devi...

Страница 11: ...atient injury could occur WARNING Before you plug the control unit into an applicable power source carefully read the instructions Failure to do so could cause patient injury personal injury or equipm...

Страница 12: ...drop the control unit Equipment damage could occur CAUTION Do not store the Synergy Air Elite Low Airloss Alternating Therapy System in direct sunlight or extreme cold conditions Equipment damage coul...

Страница 13: ...Press the Auto Firm control The Auto Firm indicator comes on and the mattress starts to inflate 12 Let the mattress fully inflate NOTE In Auto Firm mode the mattress should fully inflate within 2 to...

Страница 14: ...ry third air cell fully inflates 2 Air cells 2 5 8 11 14 17 deflate 3 Every next third air cell fully inflates 4 Air cells 3 6 9 12 15 18 deflate 5 Every final third air cell fully inflates To start t...

Страница 15: ...ells and the foam pad in the mattress assembly Make sure at least 2 5 cm exists between the patient and the top of the foam pad Auto Firm Mode Auto Firm mode quickly inflates the air cells to their ma...

Страница 16: ...e Personal injury or equipment damage could occur CAUTION Do not use harsh cleansers solvents or detergents Equipment damage could occur General Cleaning We recommend that you clean the unit with dete...

Страница 17: ...h the facility s procedures for contaminated waste b Replace with a clean top coverlet or bottom encasement Wash Industrial laundry a Break wash cycle in cold water for 10 minutes b Main wash cycle I...

Страница 18: ...ied on the manufacturer s label Maintenance WARNING Only Hill Rom authorized persons should service the Synergy Air Elite Low Airloss Alternating Therapy System Service by unauthorized persons could c...

Страница 19: ...thers with Normal Air Pressure 1 Put the unit in Static mode Unzip the top coverlet from the bottom encasement and examine the air tubes between the air cells and the main air supply line for kinks Ad...

Страница 20: ...torage and Handling CAUTION Do not store the Synergy Air Elite Low Airloss Alternating Therapy System in direct sunlight or extreme cold conditions Equipment damage could occur Feature Specification S...

Страница 21: ......

Страница 22: ...1 0 347 32 35 00 Nordic Region Hill Rom AB Ekbacksv gen 20 168 69 Bromma Sweden Tel 46 0 8 564 353 60 Fax 46 0 8 564 353 61 se marketing hill rom com sterreich Hill Rom Austria GmbH Bueropark Donau Ha...

Отзывы: