![Hill-Rom careselections Synergy Air Elite Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/hill-rom/careselections-synergy-air-elite/careselections-synergy-air-elite_user-manual_2134138014.webp)
8
Synergy® Air Elite Low Airloss Alternating Therapy System User Manual
(139122 REV 1)
Instructions for Use
CPR Mode
To quickly deflate the mattress for CPR
procedures, pull the
CPR
release on the side
of the control unit to disconnect the hose
connector.
Alternating Therapy Mode
Alternating Therapy mode inflates and deflates the 18
air cells in this order:
1. Every third air cell fully inflates
2. Air cells 2, 5, 8, 11, 14, 17 deflate
3. Every next third air cell fully inflates
4. Air cells 3, 6, 9, 12, 15, 18 deflate
5. Every final third air cell fully inflates
To start the Alternating Therapy mode, do as follows:
1. Press the
Alternate
control. The
Alternate
indicator comes on.
2. To set the necessary cycle time interval of the
therapy, press the
Cycle Timer
control to adjust the
Cycle Timer
setting in 5 minute increments
between 5 minutes and 95 minutes.
NOTE:
To estimate the total cycle time, multiply the
Cycle
Timer
setting by
4
.