47
TRIGger INPUT TERMination
Command:
TRIG:INPUT:TERM
imp
<CR><LF>
Response:
OK<CR><LF>
Query:
TRIG:INPUT:TERM?<CR><LF>
Response:
imp
<CR><LF>
Description: Sets the trigger input termination impedance. The possible values
for
imp
are shown below.
50OHM
50
Ω
internal termination
HIGHZ
High impedance internal termination
Example:
Set the internal trigger termination to 50 Ohm.
TRIG:INPUT:TERM 50OHM<CR><LF>
reply:
OK<CR><LF>
Example:
Query the unit for the trigger termination source.
TRIG:INPUT:TERM?<CR><LF>
reply:
50OHM<CR><LF>
TRIGger INPUT VOLTage
Command:
TRIG:INPUT:VOLT
volts
<CR><LF>
Response:
OK<CR><LF>
Query:
TRIG:INPUT:VOLT?<CR><LF>
Response:
volts
<CR><LF>
Description: Sets the trigger voltage level.
volts
can be set from –2.40 to +4.60 in
0.10 increments. It is not necessary to include the sign for zero or positive
values. The query response will always include the sign.
Example:
Set the input trigger level to +2.50V.
TRIG:INPUT:VOLT 2.5<CR><LF>
reply:
OK<CR><LF>
Example:
Query the input trigger level.
TRIG:INPUT:VOLT?<CR><LF>
reply:
+ 2.5<CR><LF>