![High Voltage DTS-60A Скачать руководство пользователя страница 64](http://html.mh-extra.com/html/high-voltage/dts-60a/dts-60a_operators-manual_2130381064.webp)
DTS-60A O
PERATORS
M
ANUAL
54
5.2.1 Select a Test
The Select a Test menu displays the list of all the existed tests. These
are the Simple Test, the Standards and the User Defined Tests, Figure
4-3. An arrow at the right side of a test's name indicates that this is the
last selected test. The latter is the test that will be started after turning-
on the equipment.
Figure 4-3 –
Select Program Menu
SELECT TEST
IEC 156/95 <--
VDE 0370
IEC 156/63
ASTM D1816/84A
ASTM D877/87
TEST
CREATE
SELECT
EDIT
DELETE
STEP 1
From the TEST menu point to SELECT.
Then press the ENTER key.
STEP 2
The system displays all the available tests. The
default test is indicated with an arrow at the left
side.
Use the UP and DOWN keys to select a test for
execution.
UP
DOWN
SELECT TEST
IEC 156/95 <--
VDE 0370
IEC 156/63
ASTM D1816/84A
STEP 3
Press the ENTER key to execute the selected test.
From now on this test will be the default test.
ENTER
12:10 01/08/05 65 °F
12:10 01/08/05 65 °F
12:10 01/08/05 65 °F
ASTM D877/87
ENTER
Содержание DTS-60A
Страница 1: ...DTS 60A Dielectric Oil Test Set Operator s Manual Version 1 20 HighVoltage INC ...
Страница 3: ...DTS 60A OPERATORS MANUAL ...
Страница 5: ...DTS 60A OPERATORS MANUAL ...
Страница 7: ...DTS 60A OPERATORS MANUAL ...
Страница 10: ......
Страница 48: ...DTS 60A OPERATORS MANUAL 38 ...
Страница 56: ...DTS 60A OPERATORS MANUAL 46 Figure 4 11 Test Program Execution Example ...
Страница 57: ...DTS 60A OPERATORS MANUAL 47 ...
Страница 77: ...DTS 60A OPERATORS MANUAL 67 prg 0 stage 27 003 0 kV s breakdown 00 00 00 00 10 00 00 10 00 end store prg 0 end ...
Страница 81: ...DTS 60A OPERATORS MANUAL 71 ...
Страница 87: ...DTS 60A OPERATORS MANUAL 77 ...
Страница 106: ...DTS 60A OPERATORS MANUAL 96 Figure 8 17 Reading specific result id Figure 8 18 history erase Command ...
Страница 111: ...DTS 60A OPERATORS MANUAL 101 ...